BENNING CM 8
16
- a number of digits (i.e. counting steps of the last digit).
This measuring accuracy applies to temperatures from 18 °C to 28 °C and a
relative air humidity lower than 80 %.
7.1 DC voltage ranges
The input resistance is 3 MΩ.
Measuring
range
Resolution
Measuring
accuracy
Overload
protection
60 V
0.01 V
± (0.7 % of the measured value + 5 digits)
600 V
eff
600 V
0.1 V
± (0.7 % of the measured value + 5 digits)
600 V
eff
7.2 AC voltage ranges
The input resistance is 3 MΩ in parallel 100 pF.
Measuring
range
Resolution
Measuring accuracy *
1
within the
frequency range 45 Hz – 500 Hz
Overload
protection
60 V
0.01 V
± (1 % of the measured value + 5 digits)
600 V
eff
600 V
0.1 V
± (1 % of the measured value + 5 digits)
600 V
eff
*1
The measured value is obtained and displayed as real r.m.s. value (True
RMC, AC coupling). The measuring accuracy is specified for a sinusoidal
curve.
In case of non-sinusoidal curves, the accuracy of the displayed value de-
creases. There is no additional error if the crest factor is within following
specification (AC V, 45 Hz - 65
Hz):
Range:
Crest
factor:
0 ~ 250 V
up to 3
250 V ~ 600 V from 3 (250 V) linear decreasing to 1.42 (600 V
eff
)
Peak Hold: Peak Max/ Peak Min (AC V)
Measuring range
Measuring accuracy
85.0 V... 850 V
± (3 % of the measured value +15 digits)
7.3 DC current ranges
Measuring
range
Resolution
Measuring accuracy
Overload
protection
600 A
0.1 A
± (1.5 % of the measured value + 5 digits)
600 A
eff
The specified accuracy for conductors that are centrally clamped by means of
the current clamp
M
(see figure 4 DC/ AC current measurement). For conduc-
tors that are not centrally clamped, an additional error of 1 % of the displayed
value has to be considered.
Maximum remanence error: 1 % (in case of repeated measuring)
Condition: Null balance before measuring!
7.4 AC current ranges
Measuring
range
Resolution
Measuring accuracy
*1
within the
frequency range 45 Hz – 65 Hz
Overload
protection
600 A
0.1 A
± (1.5 % of the measured value + 5 digits)
600 A
eff
within the frequency range
66 Hz - 400 Hz
600 A
0.1 A
± (2.5 % of the measured value + 5 digits)
600 A
eff
*1
The measured value is obtained and displayed as real r.m.s. value (True
RMC, AC coupling). The measuring accuracy is specified for a sinusoidal
curve.
In case of non-sinusoidal curves, the accuracy of the displayed value de-
creases. Thus, an additional error results for
Crest factor from 1.4 to 2.0 additional error + 1 %
Crest factor from 2.0 to 3.0 additional error + 2.0 %
within following specification: (AC A, 45 Hz - 65 Hz)
Range:
Crest
factor:
0 ~ 250 A
up to 3
250 A ~ 600 A from 3 (250 A) linear decreasing to 1.42 (600 A)
The specified accuracy for conductors that are centrally clamped by means of
the current clamp
M
(see figure 4 DC/ AC current measurement). For conduc-
tors that are not centrally clamped, an additional error of 1 % of the displayed
value has to be considered.