BeneVision TM80 Telemetry Monitor Operator’s Manual
10 - 3
with a new telemetry monitor, you must select
No
. Otherwise, chaotic patient
data may occur.
◆
If neither
Yes
nor
No
is selected, the wireless setup confirmation message will
disappear in three minutes. Besides, the “Pairing failure” message will be
displayed in the prompt message area of the telemetry monitor. The “MPAN
Disconnected” message is displayed in the message area of the BP10. In this
case, follow steps in
10.2 Pairing a Telemetry Monitor with a BP10
telemetry monitor with the BP10 again.
■
When NIBP measurement mode is set to
Manual
,
Auto
, or
Sequence
and the BP10 is
paired successfully with the telemetry monitor, the wireless setup confirmation
message “Select Yes to continue Patient in BP10. Select No to continue Patient in
Telemetry.” will not be displayed. Patient-related settings from the telemetry monitor
will be transferred to the BP10. But NIBP-related settings displayed on the main
screen of the telemetry monitor will come from the BP10. The patient monitored by
BP10 will be discharged automatically and the data history related to this patient will
be cleared.
■
Patient Category can be changed at the telemetry monitor only.
■
A patient cannot be discharged from the BP10 and NIBP-related settings are grayed
out at BP10. To discharge a patient from BP10, you need to unpair the BP10 form the
telemetry monitor by following steps in
10.3 Unpairing the Telemetry Monitor with
.
10.3 Unpairing the Telemetry Monitor with the BP10
You can unpair the telemetry monitor with the BP10 either via the telemetry monitor or via
the BP10.
10.3.1 Unpairing via the Telemetry Monitor
You can choose either of the two methods to unpair the telemetry monitor with the BP10
via the telemetry monitor.
Option 1
:
1.
Press the
key to enter the main menu of the telemetry monitor.
2.
Tap
Wireless Modules
.
3.
Select the device you wish to disconnect from the telemetry monitor from the
devices list under
Devices
.
4.
Tap
Disconnect
. The connected device is disconnected from the telemetry monitor.
When the telemetry monitor is unpaired with the BP10 successfully:
■
The status of the BP10 unpaired will be displayed as
Disconnect
on the list of
Devices
at the telemetry monitor.
■
The
icon will be changed to
in the top-right corner of the telemetry
monitor’s screen and of the BP10.
Содержание TM80
Страница 1: ...BeneVision TM80 Telemetry Monitor Operator s Manual...
Страница 2: ......
Страница 20: ...1 6 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Страница 29: ...BeneVision TM80 Telemetry Monitor Operator s Manual 2 9...
Страница 30: ...2 10 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Страница 46: ...3 16 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Страница 50: ...4 4 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Страница 56: ...5 6 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Страница 66: ...6 10 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Страница 135: ...BeneVision TM80 Telemetry Monitor Operator s Manual 11 7 If you wish to return to the event list tap the icon...
Страница 136: ...11 8 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Страница 174: ...14 8 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Страница 192: ...15 18 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Страница 200: ...16 8 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Страница 206: ...17 6 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Страница 238: ...D 4 BeneVision TM80 Telemetry Monitor Operator s Manual This page intentionally left blank...
Страница 239: ......
Страница 240: ...P N 046 012662 00 6 0...