Chapter 10 Limitations
Please observe the following limitations for obtaining accurate results of
BeneCheck Blood Glucose Monitoring System :
1. For In-vitro diagnostic use only ( External use only ).
2. Do not dispose test strips in high humidity or direct sunlight. Don’t freeze the
strips.
3. BeneCheck Blood Glucose Test Strip is for single use only, please do not reuse
the test strip.
4. Handle the strip with clean and dry hands; lancing site for the test should be
cleaned and dried thoroughly.
5. Do not perform the test while a cell phone is used beside; prevent the
electronic signal interference.
6. Oxygen concentration low in air may cause the test inaccurate; there is no
effect from altitude up to 10,000 feet above sea level.
7. Medical therapy for intravenous injection, such as immunoglobulin
p r e p a ra t i o n s a n d p e r i t o n e a l d i a l y s i s s o l u t i o n s , c o n t a i n i n g s u g a r
compounds may cause higher test results.
8. Hematocrit (Hct) range from 30-55% is acceptable. Hct less than 30%
may cause elevated results. Hct above 55% may cause reduced glucose
results versus a whole blood reference.