9.2 Limitations of Glucose Test
BeneCheck Glucose Test Strips give accurate results when the following limitations
are observed :
●
Hematocrit (Hct) percentage may affect test result. If Hct levels less than 30%
may cause incorrect high measurement results and Hct levels greater than 55%
may cause incorrect low measurement results. Please consult your medical
professionals if you do not know your Hct level.
●
The test strips react with only D-glucose and do not react with other sugars
which may be present in blood.
●
Use only fresh capillary finger whole blood. Do not use serum or plasma.
●
BeneCheck Glucose Test Strips may be used at altitude up to 10,000 feet
without an effect on test results.
Additional Information for Healthcare Professionals :
●
When venous whole blood is used for the test, it can be collected into
Содержание Chol.GLU