1 2 1 - 1 8 1
3 . 1 - 4 . 7
After count down and the screen will display the control test
results. Compare the results to the control range listed on
the test strip vial.
Note :
●
Do not reuse test strips.
●
Newly opened vials of control solution and test strips must be marked with the
opening date.
●
Close the test strip vial cap immediately after taking out the strip.
● Do not use BeneCheck CHOL.GLU Dual Monitoring System to do any test until you
can get the control test result falls within the range listed on the test strip vial.
●
If the test results fall outside the control range, please refer to Chapter 7 for
trouble shooting.