background image

Multi-Monitoring Meter

Tot

al  

Ch

ol

es

te

ro

l  

Bl

ood Glucos

e  U

ric

 A

cid

Check Your Health

Hand in Hand,  lth

Hand in Hand, 

User’s Manual

Total Cholesterol

Glucose

Uric Acid

General Life Biotechnology Co., Ltd.

5F, No.240, Shinshu Rd., Shin Juang Dist., 

New Taipei City, 242 Taiwan

www.BeneCheck.com.tw

MDSS GmbH

Schiffgraben 41

30175 Hannover, Germany

TEL : +886-2-22053698

FAX : +886-2-22080779

E-mail : [email protected]

Doc.#:FIG-312M-C2-101-2 Jan, 2017

BK3-12M-D

Model:BK3-12M-D

Содержание BK3-12M-D

Страница 1: ... s Manual Total Cholesterol Glucose Uric Acid General Life Biotechnology Co Ltd 5F No 240 Shinshu Rd Shin Juang Dist New Taipei City 242 Taiwan www BeneCheck com tw MDSS GmbH Schiffgraben 41 30175 Hannover Germany TEL 886 2 22053698 FAX 886 2 22080779 E mail service glbiotech com tw Doc FIG 312M C2 101 2 Jan 2017 BK3 12M D Model BK3 12M D ...

Страница 2: ...p and Code Strip 6 Installing Replacing the Batteries 7 Set the Date and Time 7 8 Code the Meter 8 Before Testing 9 Perform the Test 10 12 Alternate Site Testing AST 12 14 Care after Testing 15 16 3 4 Chapter 4 Meter Memory Function 17 18 Chapter 5 Control Solution Test 19 20 Chapter 7 Care and Maintenance Chapter 8 Error Message and Trouble Shooting 23 24 Chapter 9 Specification 25 7 1 7 2 Storin...

Страница 3: ...ethods Test Glucose Total Cholesterol Uric Acid Standard NIST SRM 917c NIST SRM 911b NIST SRM 913a Method Glucose Oxidase Abell Kendall Uricase UV Please carefully read the instructions and test procedures in this user s manual before using Users should consult healthcare professional before making any important medical decision Description Quantity BeneCheck Multi Monitoring Meter Model BK3 12M D...

Страница 4: ...t meter Battery Lid Slide to open 3 4 Phone Jack No function 1 3 Labelling and Information Do not reuse Keep dry Read instructions Use by EU Representative Manufactured by Keep away from sunlight Caution consult accompanying documents Operation temperature limitation Operation In vitro diagnostic Lot number Storage Transportation 4 Storage Transportation Condition CE certification Serial number 90...

Страница 5: ...rip Loading Icon Test Mode Icon Memory Mode Icon Test Strip GLU Glucose CHOL Total Cholesterol UA Uric Acid Code Strip The code strip is used to code the meter Please do the coding step when you use a new vial of CHOL or UA strip Contact Bar Insert this end into the test port of the meter UA CHOL Test Mode Code 5 6 7142 Electronic Contact Bars Insert this end into meter test port Sample Inlet The ...

Страница 6: ...e one unit Click S button to enter next setting The meter will turn off automatically after setting Note Adjust the correct time and date is important to help managing your health records 2 3 Code the Meter Each vial of CHOL or UA strip has a code strip to code your meter Code your meter when you first use it or open a new vial of strips With the Auto Strip Recognition function once you code your ...

Страница 7: ...he strip contact bar into meter test port The meter will turn on automatically 5 6 Make sure the code numbers are the same on screen and strip vial label in CHOL or UA test You can start testing after appear Pull back the sliding barrel till it clicks and the lancing device is ready to use 7 8 Hold the lancing device firmly against sampling site Press the release button 9 10 Lancet Holder Ejection...

Страница 8: ...ay result in inaccurate test result or damage the system Please finish the test within 5 minutes or the meter will turn off automatically The meter will not turn on if you insert the wrong end or wrong side of strip Do not use expired strip Warning Lancets cannot be reused Please always use certified lancets to ensure safety 3 3 Alternate Site Testing AST The meter allows you to test glucose from ...

Страница 9: ... Choose an area away from bone visible veins and hair Please massage or warm the site to increase blood flow in order to get sufficient amount of blood sample Lancing and sampling from an alternative site 1 Repeat the steps1 7 in Chapter 3 2 Replace the lacing device tip with adjustable AST tip 2 Then press and hold the lancing device against your palm or forearm for a few seconds then press the r...

Страница 10: ...dL 2 59 mmol L 400 mg dL 10 35 mmol L Note Please check your meter performance regularly Dropping bumping or other violent impact may damage the meter or cause malfunction Do not use the meter in an environment with possible magnetic electromagnetic and radioactive interferences Choking hazard keep away from children Warning Do not disassemble the meter for any reason Do not modify this equipment ...

Страница 11: ...ote The control results are not included in the average GLU control Record CHOL or UA control Record 1 No test strip in the meter 2 Press M button for 3 seconds till a short beep to enter the memory mode 3 After a full display on screen press M button to switch between GLUC UA and CHOL mode Press S button to select 4 In GLU memory mode it will display 7 14 21 28 days average first then individual ...

Страница 12: ...l of control solution and test strips must be marked with the opening date Close the strip vial lid completely after use Do not use the meter if the control range do not fall in the correct range If the control test keeps result wrongly please contact an authorized distributor for help and service Control solutions are not included Please contact an authorized distributor for checking the meter ki...

Страница 13: ...ease follow the instructions and check Chapter 8 for specifications Please avoid bump or other violent behavior it may caused damage to the meter Do not use this meter in a very dry environment especially if synthetic materials are present It may cause damaging static discharges in a dry environment Do not use the meter near the source of electromagnetic radiation ex electrical or electronically e...

Страница 14: ...sists please contact the distributor for service 23 24 Problem with strip Review the instructions and try again with the same test strip If the problem persists please contact the distributor for service Uncompleted test caused by removing the test strip during the measurement Review the instructions and try again with a new test strip Please don t remove the strip before the test is completed Use...

Страница 15: ...equirements as per EN61326 The purpose of these requirements is to ensure that meter does not affect or being affected by other devices during operation Warning Use of this equipment adjacent to or stacked with other equipment should be avoided because it could result in improper operation If such use is necessary this equipment and the other equipment should be observed to verify that they are op...

Страница 16: ... accordance with the instructions may cause harmful interference to radio communications However there is no guarantee that interference will not occur in a particular installation If this equipment does cause harmful interference to radio or television reception which can be determined by turning the equipment off and on the user is encouraged to try to correct the interference by one or more of ...

Отзывы: