2.1   The fundamental functions
The LIM function is to calculate the max. value of the total hazard 
current that would flow through a solid connection between the 
isolated current carrying conductor associated with the higher 
impedance and ground. The total hazard current is displayed 
on a 7-segment LED display and an analog bargraph with 16 
LEDs. The level will increase as additional loads are connected 
to the system or when a ground fault is slowly developing or has 
occured.
The bargraph has two scales, one for 2mA and one for 5 mA alarm 
setting. The two yellow LEDs on the left near the bargraph 
indicate, which alarm setting is selected and which scale is valid.
There is a visual and audible alarm when the total hazard current 
exceeds the setting of either 2 or 5mA. The visual alarm remains 
during duration of the fault, the buzzer can be muted at any time. 
When muted, the yellow LED inside the mute button comes on to 
indicate a muted condition.
Activating the test button starts the following procedures:
-    Audible alarm check
-    Illumination of the analog LED bargraph
-    Illumination of the 7-segment LED displays of the total hazard 

current and the optional transformer load 

-    Internal hardware check and self calibration by using internal 

faults (the numeric Display of the optional transformer load 
indicates CH or CL depend on 120 or 240V).

If these tests are successful, the LIM goes in the operating 
condition, otherwise the LIM displays an errorcode. After the fault 
of the error has abolished it works again in the operating condition, 
when the internal tests are finished. 
The LIM also performs an internal hardware check and a 
self-calibration after 3 hour and then every 12 hours. If this checks 
aren´t successful the LIM displayes an error-code too.
The LIM has provisions for connecting one or two remote 
indicators with or without meter or digital display. Relay output 
contacts are also available which can be wired into a circuit to 
trigger an external alarm.

2. General Information

10

11

Содержание LIM2000plus LIM2000-1

Страница 1: ...TGH 1337 Edition 01 06 Operating Manual LIM2000plus Line Isolation Monitor LIM2000 1 LIM2000 1CB LIM2000 3CB...

Страница 2: ...r LIM2000 1 LIM2000 1CB and LIM2000 3CB 6 1 6 Inspection transport and storage 7 1 7 Important 7 1 8 Explanation of symbols and notes 8 1 9 Directives for installation 9 2 General Information 11 2 1 T...

Страница 3: ...ing arising therefrom Correct use also includes compliance with all instructions from the operating manual adherence to any inspection intervals As a basic principle our General conditions of Sale and...

Страница 4: ...evertheless when it is being used hazards may occur to the life and limb of the user or of third parties or there may be adverse effects on the LIM or on other valuable property The LIM must only be u...

Страница 5: ...means a possible dangerous situation Failure to comply with these warnings means that slight physical injury or damage to property may ensue if the relevant precautions are not taken This symbol gives...

Страница 6: ...he mute button comes on to indicate a muted condition Activating the test button starts the following procedures Audible alarm check Illumination of the analog LED bargraph Illumination of the 7 segme...

Страница 7: ...nsformer current After setting the parameters they are stored and after leaving the menu the LIM performs a new self test 12 13 2 General Information 2 General Information 1 Amber THC setting LEDs ind...

Страница 8: ...u and store new parameters menu END TEST 2 4 Error Codes Listing ER 1 0 Ground error LIM GNDor GND interrupted ER 2 0 AD converter calibration failed hardware failed error ER 3 0 Measuring circuit tes...

Страница 9: ...icator and the isolated power supply are made via this cable assembly designed for integration into panels headwalls and other equipment The terminal strip works best with 18 ga wire The multiconducto...

Страница 10: ...harness has a length of 20 2 5 1 Remote Indicator The MK2000 Remote Indicator series permanently mounts into standard one to xxx gang electrical boxes Low voltage wiring connects them via the connecto...

Страница 11: ...s based on the superimposition of a measuring signal on the isolated power system The signal is very small and near the frequency of the power line making it compatible with all electrical equipment d...

Страница 12: ...DC 200mA contact Rated contact voltage 250 V AC 24VDC Make capacity 4A AC 4A DC Break capacity at 24 V DC and L R 0 4 A Switching life 220 V AC 60 Hz 2 x 105 cycle Operation mode continuous LIM overl...

Отзывы: