Settings
Settings
isoHR685W-x-I-B_D00261_05_M_XXEN/12.2021
37
9.2
(5.3.4) Modbus/TCP
Settings for communication with other devices via Modbus TCP.
9.2
(5.3.4.1) Port 502
Choose whether Modbus TCP should be used:
9.2
(5.3.5) RS-485
Set the parameters for communication with other devices via the Bender sensor bus.
9.2
(5.3.5.1) BS bus
9.2
(5.3.5.2) isoData
9.2
(5.3.6) Modbus RTU
9.2
(5.4) Display
Adjust the display brightness for the ISOMETER® in the "Display" menu:
9.2
(5.4.1) Brightness
Adjust the display brightness between 0 % and 100 % in steps of 10.
If no button is pressed on the display for 15 minutes, the brightness of the display is re-
duced. If now a button is pressed, the normal brightness is restored.
•on
Modbus TCP can be used for communication with other devices.
•off
Modbus TCP cannot be used for communication with other devices.
1. Mode:
Selecting an RS-485 protocol
2. BS bus
BS bus settings
3. isoData
isoData settings
4. Modbus RTU
Modbus RTU settings (Remote Terminal Unit)
1. Address
Address adjustable from 1…90
1. Protocol
Mode adjustable 1, 2 or 3
1. Address
Entering an address from 1…247
2. Baud rate
Setting 9.6 | 19.2 | 37.4 | 57.6 | 115 kbaud
3. Parity
Setting "even" | "odd" | "none"
4. Stop bits
Setting "1" | "2" | "off"
9.2
(5.4.2) Automatic dimming
9.2
(5.5) Password
Use the password function to protect the device parameters against unauthorised adjust-
ment. The default password is 0000.
9.2
(5.5.1) Password
Enter an individual four-digit password.
9.2
(5.5.2) Status
Decide if the password query should be used:
9.2
(5.6) Commissioning
In the "Commissioning" menu you can open the ISOMETER®'s commissioning wizard
again. Pressing the commissioning button immediately starts the commissioning wizard.
After going through all the queried values, the new values are accepted by the device.
Press "ESC" to abort the process.
9.2
(5.7) Data backup
In the "Data backup" menu device settings can be saved or device settings already saved
can be restored.
9.2
(5.8) Activation
Activation of special customer profiles by Bender. The device is first configured by the
Bender service and the configuration is saved in a service profile. Activating this profile
causes a warning message. The customer can activate it as a customer-specific profile by
entering a
Service Profile PIN
. The warning message is then eliminated.
• on
Background lighting, POWER LED and button lighting are switched off after 3 min-
utes without operation and are only switched on again when any button is pressed.
Alarm LEDs light in the event of an alarm.
• off
•on
Password query active
•off
Password query inactive
•Save
The ISOMETER® saves your device settings.
•Restore
The ISOMETER® restores your initial or your saved device settings.
1. Profile:
Enter a 4-digit Service Profile PIN