Settings
Settings
isoHR685W-x-I-B_D00261_05_M_XXEN/12.2021
33
9.2
(1.12.3) Digital 1
The following parameters can be set for each of the digital outputs:
9.2
(1.12.3.1) TEST
The functional test of the digital output can be activated or deactivated. This only applies
to the manual test and not to the cyclic device self test:
9.2
(1.12.3.2) Mode
The following settings can be used to set the operating mode for the digital output:
9.2
(1.12.3.3) Function 1
Refer to
chapter 9.2 (1.12.1.3) “Function 1” on page 32
.
•on
The manual test changes the status of the digital output.
•off
The manual test does not change the status of the digital output.
•Active
In active mode, +24 V are applied internally to output Qx.
•Passive
In passive mode, ≤ 32 V are connected externally (see technical data). The
output switches the applied potential to ground.
Maximalen Ausgangsstrom beachten!
Maximaler Ausgangsstrom bei interner Spannungsversorgung
über A1/+ und A2/-: 200 mA in Summe an X1.
Beachten Sie außerdem die Formel zur Berechnung von I
LmaxX1
in den
Technischen Daten.
X1
iso685
Q
x
+
X1
iso685
Q
x
+
or
X1
iso685
Q
x
+
X1
iso685
Q
x
+
intern
extern
or
9.2
(1.12.3.4) Function 2
Refer to
chapter 9.2 (1.12.1.3) “Function 1” on page 32
.
9.2
(1.12.3.5) Function 3
Refer to
chapter 9.2 (1.12.1.3) “Function 1” on page 32
.
9.2
(1.12.4) Digital 2
Refer to
chapter 9.2 (1.12.3) “Digital 1” on page 33
.
9.2
(1.12.5) Buzzer
The following parameters can be set for the buzzer.
9.2
(1.12.5.1) TEST
The functional test of the buzzer can be activated or deactivated. This only applies to the
manual test and not to the cyclic device self test:
9.2
(1.12.5.2) Function 1
Refer to
chapter 9.2 (1.12.1.3) “Function 1” on page 32
.
9.2
(1.12.5.3) Function 2
Refer to
chapter 9.2 (1.12.1.3) “Function 1” on page 32
.
9.2
(1.12.5.4) Function 3
Refer to
chapter 9.2 (1.12.1.3) “Function 1” on page 32
.
9.2
(1.12.6) Analogue
The following parameters can be set for the analogue output.
9.2
(1.12.6.1) Mode
The following values can be set for the operating mode of the analogue output
:
•on
The manual test activates the buzzer sound.
•off
The manual test does not activate the buzzer sound.
Current output
•0-20 mA
Permissible load ≤ 600 Ω
•4-20 mA
Permissible load ≤ 600 Ω
•0-400 μA
Permissible load ≤ 4 kΩ
Voltage output
•0-10 V
Permissible load ≥ 1 kΩ
•2-10 V
Permissible load ≥ 1 kΩ
M
+
X1
X1
A
M
+
X1
X1
V