4-3 Daily Check-up of functions by the end user

The end-user is responsible for daily check-up of the equipment. In the case end user does not do this, it is

allowed that you outsource it to a technician authorized by us. 

In order to assure safety use of this equipment, please carry out the following check-up in a frequency as 

indicated below.

No. Item

(Frequency)

How to check and diagnose

Influence of not doing 

check-up

Actions when any incom-

pliance is found

1

Safety functions

(Daily. Before 

opening office)

During pre-set movement or 

auto-return movement, press any 

key for chair operation and the 

chair  should automatically stop.

Injury caused by move-

ment of the chair during 

treatment or pinching with 

chair and other equipment 

or furniture.

Call to a technician of our 

authorized dealer.

2

Function of each 

switch

(Daily. Before 

opening office)

Each switch for chair operation 

should work correctly. 

Defect of movement may 

cause an accident.

Call to a technician of our 

authorized dealer.

3

Wiping off water

(Daily. Before

closing office)  

Make sure that no water is left 

that were spilt during treatment.

Water will cause rust

Wipe off water with a soft 

and dry cloth.

4

Stain removal of 

external parts

(Dail y. Before 

closing office)

Visually make sure that no dust, 

chemical or other material is left 

on the chair.

Such chemicals will cause 

discoloration, dete-

rioration and break of the 

plastics.

Thoroughly clean the

chair following the

cleaning instruction in 

this book, regardless the 

result of this check-up. 

For tough stains, use mild 

detergent solvency.  

5

Main power 

switch

(Daily. Before 

closing office)

Make sure that the main power 

switch is off

Accident or defect of the 

equipment

In the case the main 

power cannot be turned 

off, Call to a technician of 

our authorized dealer.

6

Moving parts

(Weekly)

Make sure that the chair does not 

make any abnormal noise during 

automatic movement

Accidend or defect of the 

equipment

Call to a technician of our 

authorized dealer.

7

Oil leakage

(Weekly)

Make sure that there is no hy-

draulic oil on the floor or chair 

base leaking from the cylinder. 

Defect of the equipment

Call to a technician of our 

authorized dealer.

8

Power supply 

cable

(Monthly)

Power supply cable should not 

be abnormally bent, pintched or 

damaged, and the plug should 

befirmly inserted to the power 

outlet and not covered with dust.

Accident or defect of the 

equipment

Clean and rearrange the 

cable. For replacement, 

call to a technician of our 

authorized dealer.

7

Содержание X-Calibur BEL-20N

Страница 1: ...structions for the BEL 20N Chair The instructions contained in this booklet should be thoroughly read and understood before operating the chair After the installation has been completed keep this manu...

Страница 2: ......

Страница 3: ...r Switch Applied part complying with the specified requirements of IEC60601 1 2005 to provide protection against electrical shock particularly regarding allowable patient leakage current and patient a...

Страница 4: ...8 15 0 to 70 above horizontal Backrest Synchronized Tilting 10 to 20 30 Left 30 Right of the center line Either armrest can be rotated outward 1 pre set and auto return AC120V 60Hz 3 7A 10A 125V Curr...

Страница 5: ...llowed Do not allow patient to sit on the headrest armrest backrest or the end of seat Do not give too much load on the chair cushions Failure to do so might lead to malfunction or an accident such as...

Страница 6: ...mentarily depress 0 button on the foot control the chair will return to the initial position The seat is fully lowered and the backrest is in the upright position c Last Position Memory Momentarily de...

Страница 7: ...d headrest cushion while pressing latch release button 4 Tilt the headrest cushion until the desired upper arch position is obtained Headrest Release Lever 2 Chair Rotation To rotate the chair release...

Страница 8: ...mbinations Foam spray products Benzalconium chloride solutions Use with Caution Water based phenolic disinfectants following manufacturer s exact instructions for use Warranty does not cover damage to...

Страница 9: ...water with a soft and dry cloth 4 Stain removal of external parts Dail y Before closing office Visually make sure that no dust chemical or other material is left on the chair Such chemicals will cause...

Страница 10: ...nfecious device but prior to disposal make sure that no parts of the device is infected Follow the federal state and local regulations for disposal 4 5 Replacement of consumables Replacement must be d...

Страница 11: ......

Страница 12: ...6203 TAKARA BELMONT CORPORATION Manufacturer 1 1 2 Chome Higashi shinsaibashi Chuo ku Osaka 542 0083 Japan TEL 81 6 6213 5945 Fax 81 6 6212 3680 This equipment should installed by an professional tec...

Отзывы: