3-4 Positioning of Headrest

1. Seat patient upright and all of the way back on the chair.

2. Position the bottom of the headrest cushion so it is even

    with the occipital of the patient’s head.

Viewing the lower arch

1. Using the latch release button (white), pivot the headrest cush-

    ion halfway toward the patient’s head.

2. Have the patient lean back on the headrest.

3. Using the chair controls, place the patient in the desired operat-

    ing position.

Viewing the upper arch

1. Lower the backrest and raise the base of the chair.

2. Slide the headrest in.

3. Support the patients head/headrest cushion while pressing latch release button.

4. Tilt the headrest cushion until the desired upper arch position is obtained.

Headrest Release

Lever

2) Chair Rotation

   To rotate the chair, release lock lever on right side of the chair, by pushing downward. Pull lever to lock. 

3) Cancellation

a. Safety plate is provided below base link cover to stop motion of chair when object is accidentally caught 

    underneath.

b. Automatic motion activated either by automatic return switch or preset switch, can be cancelled at any time 

by depressing any control switches. 

4) Headrest Adjustment

a. Press down or pull up the headrest for height adjustment.

b. Grab the headrest release lever on headrest mechanism.

T

ake care not to pintch the 

 ngers with the headrest mechanism.

CAUTION

5

Содержание X-Calibur BEL-20N

Страница 1: ...structions for the BEL 20N Chair The instructions contained in this booklet should be thoroughly read and understood before operating the chair After the installation has been completed keep this manu...

Страница 2: ......

Страница 3: ...r Switch Applied part complying with the specified requirements of IEC60601 1 2005 to provide protection against electrical shock particularly regarding allowable patient leakage current and patient a...

Страница 4: ...8 15 0 to 70 above horizontal Backrest Synchronized Tilting 10 to 20 30 Left 30 Right of the center line Either armrest can be rotated outward 1 pre set and auto return AC120V 60Hz 3 7A 10A 125V Curr...

Страница 5: ...llowed Do not allow patient to sit on the headrest armrest backrest or the end of seat Do not give too much load on the chair cushions Failure to do so might lead to malfunction or an accident such as...

Страница 6: ...mentarily depress 0 button on the foot control the chair will return to the initial position The seat is fully lowered and the backrest is in the upright position c Last Position Memory Momentarily de...

Страница 7: ...d headrest cushion while pressing latch release button 4 Tilt the headrest cushion until the desired upper arch position is obtained Headrest Release Lever 2 Chair Rotation To rotate the chair release...

Страница 8: ...mbinations Foam spray products Benzalconium chloride solutions Use with Caution Water based phenolic disinfectants following manufacturer s exact instructions for use Warranty does not cover damage to...

Страница 9: ...water with a soft and dry cloth 4 Stain removal of external parts Dail y Before closing office Visually make sure that no dust chemical or other material is left on the chair Such chemicals will cause...

Страница 10: ...nfecious device but prior to disposal make sure that no parts of the device is infected Follow the federal state and local regulations for disposal 4 5 Replacement of consumables Replacement must be d...

Страница 11: ......

Страница 12: ...6203 TAKARA BELMONT CORPORATION Manufacturer 1 1 2 Chome Higashi shinsaibashi Chuo ku Osaka 542 0083 Japan TEL 81 6 6213 5945 Fax 81 6 6212 3680 This equipment should installed by an professional tec...

Отзывы: