DIFFERENCES BETWEEN PATIENT MODES
Adult Mode
Neonatal Mode
Default Set Point
33.0°C
33.5°C
Adjustable Default
Yes (in Settings
Screen 1)
No
Adjustable Default
Options
33.0°C, 34.0°C,
35.0°C, 36.0°C
n/a
>2.0°C between Set
Point and Core
Temperature stops
water flow to the wrap
Only if patient core
temp <31.0°C
Yes, always*
Patient temp drops
below 31.0°C
No water flow
to the wrap
No water flow
to the wrap
*After the user acknowledges the message on the screen, water will once again flow to the wrap and CritiCool will not alarm for
that condition for 30 minutes. If after 30 minutes the alarm conditions apply, it will once again stop water flow and alert the user.
Содержание Criticool 6.3
Страница 1: ...701 00235 Rev B CRITICOOL 6 3 OPERATIONAL TRAINING UPDATED MAY 2020 701 00235 Rev B...
Страница 2: ...701 00235 Rev B INTRODUCTION...
Страница 5: ...2020 BELMONT MEDICAL TECHNOLOGIES Proprietary Confidential 5 701 00235 Rev B SIMPLE PRECISE EFFECTIVE...
Страница 12: ...701 00235 Rev B SETTING UP CRITICOOL...
Страница 14: ...2020 BELMONT MEDICAL TECHNOLOGIES Proprietary Confidential 14 701 00235 Rev B UNPACKING CRITICOOL...
Страница 46: ...701 00235 Rev B MESSAGES AND TROUBLESHOOTING...
Страница 62: ...701 00235 Rev B MAINTENANCE...
Страница 68: ...2020 BELMONT MEDICAL TECHNOLOGIES Proprietary Confidential 69 701 00235 Rev B PACKING CRITICOOL FOR SHIPMENT...
Страница 69: ...2020 BELMONT MEDICAL TECHNOLOGIES Proprietary Confidential 70 701 00235 Rev B PACKING CRITICOOL FOR SHIPMENT...
Страница 72: ...701 00235 Rev B THANK YOU...