![Belmont Criticool 6.3 Скачать руководство пользователя страница 29](http://html1.mh-extra.com/html/belmont/criticool-6-3/criticool-6-3_operational-training_2730288029.webp)
29
© 2020 BELMONT MEDICAL TECHNOLOGIES |
Proprietary & Confidential
Internal Use Only
701-00235 Rev. B
SECURING THE CUREWRAP AROUND THE PATIENT
1.
Confirm that CureWrap is filled with water
2. Secure CureWrap
on the patient (see CureWrap
Instructions For Use)
Note:
– Ensure clamps to the wrap are open
– Ensure CureWrap fits loosely; there should be a finger’s
width space
– Do not use the CureWrap to lift the patient
– Avoid handling the CureWrap with sharp objects
– Constant clicking sound may indicate water obstruction
○
Check to ensure the clamps are open and water is allowed to
flow to CureWrap and that water is flowing
Содержание Criticool 6.3
Страница 1: ...701 00235 Rev B CRITICOOL 6 3 OPERATIONAL TRAINING UPDATED MAY 2020 701 00235 Rev B...
Страница 2: ...701 00235 Rev B INTRODUCTION...
Страница 5: ...2020 BELMONT MEDICAL TECHNOLOGIES Proprietary Confidential 5 701 00235 Rev B SIMPLE PRECISE EFFECTIVE...
Страница 12: ...701 00235 Rev B SETTING UP CRITICOOL...
Страница 14: ...2020 BELMONT MEDICAL TECHNOLOGIES Proprietary Confidential 14 701 00235 Rev B UNPACKING CRITICOOL...
Страница 46: ...701 00235 Rev B MESSAGES AND TROUBLESHOOTING...
Страница 62: ...701 00235 Rev B MAINTENANCE...
Страница 68: ...2020 BELMONT MEDICAL TECHNOLOGIES Proprietary Confidential 69 701 00235 Rev B PACKING CRITICOOL FOR SHIPMENT...
Страница 69: ...2020 BELMONT MEDICAL TECHNOLOGIES Proprietary Confidential 70 701 00235 Rev B PACKING CRITICOOL FOR SHIPMENT...
Страница 72: ...701 00235 Rev B THANK YOU...