Installation and operating manual
EN
20 METPOINT
®
BDL compact
9.4. Terminal assignment connector "E" (RS485 - Modbus)
9.5. Terminal assignment connector "A – B" (alarm relay)
The alarm outputs are designed as floating switchover contacts. The alarm signal can be forwarded to a control desk, etc.
through the floating contacts. The connectors for the alarm contacts are labelled as "A" and "B".
DANGER!
Mains voltage
When wiring the electric power supply line, ensure that the double or reinforced insulation between
the electric circuits and the secondary circuit remains intact.
NOTICE!
Mains voltage
The additional insulation must be suitable for a test voltage of 1500 VAC. The insulation must be at
least 0.4 mm thick
e.g. insulation hose, type BIS 85 from Bierther GmbH.
The additional insulation of the power cables (mains connection, alarm and signalling relays) can be implemented as follows:
1
2
3
1
2
3
1
- Terminals (plug-type connectors)
2
- Flexible insulating tubing for the power cables
3
- Connecting cable
1 2 3 4 5 6
Common
RS
485
(B
)
RS
485
(A
)
Common
RS
485
(B
)
RS
485
(A
)
Содержание METPOINT BDL compact
Страница 1: ...METPOINT BDL compact EN english 10 089 Intelligent screen saver Installation and operating manual ...
Страница 115: ...EN Installation and operating manual METPOINT BDL compact 115 ...
Страница 116: ...Installation and operating manual EN 116 METPOINT BDL compact 15 Declaration of Conformity ...