8

6.2.3  Access code for node configuration
          menu   ‘CodE’
Access  to  the  node  configuration  menu  may  be
protected  by  a  four  digit  security  code  which  must
be  entered  to  gain  access.    New  instruments  are
supplied  configured  with  the  default  code  0000
which 

disables 

this 

protection 

and 

allows

unrestricted access to the configuration menu.

To  enter  a  new  access  code  select  ‘CodE’  in  the
configuration  menu  by  operating  the 

  or 

button.      To  enter  the  function  press  P  which  will
reveal  the  current  access  code  with  one  digit
flashing. The code may be changed using the 

 or

  button  to  adjust  the  flashing  digit  and  the  P

button to move control to the next digit.  When the
required code has been selected, press E to enter
the selection and return to the ‘CodE’ prompt in the
configuration menu.    The revised access code will
be  activated  when  the  indicator  is  returned  to  the
display mode.

If  the  access  code  is  lost  please  contact  BEKA
associates.

Fig 4 Node configuration menu

6.3  Listener configuration menu
The  listener  configuration  menu  is  separated  into
three  sections,  ‘SCAn’,  ‘LISt’  and  ‘Edit’,  plus  three
housekeeping functions ‘CodE’, ‘tYPE’ and  ‘rESEt’
as shown in Fig 6.

‘SCAn’  is  the  simplest  method  of  configuration  if
the  fieldbus  Device  Address  and  Variable  Identity
of  the  data  to  be  displayed  by  the  BA444DF-F
listener are not known.

In  the  ‘SCAn’  function  pressing  the  P  button
causes  the  BA444DF-F  indicator  to  search  the
fieldbus  segment  to  which  it  is  connected  for  all
publishers.  Fig 5 shows how the indicator displays
the  hexadecimal  Device  Address  and  Variable
Identity associated with each publisher.

A  Publisher  is  a  device  that  places  it's  output
data  onto  the  bus  as  part  of  a  cyclical  pre-
determined data exchange schedule.

The  Device  Address  is  a  unique  address  that
identifies each fieldbus instrument connected to
the bus.

The  Variable  Identity  is  the  fieldbus  device
Data  Link  Connection  End  Point  (DLCEP)
which  identifies  each  Function  Block  output
(process  variable)  transmitted  by  a  fieldbus
device 

such 

as 

flow 

and 

temperature

transmitted by a fieldbus flowmeter.

The 

  and 

  push  buttons  scroll  the  display

through all the publishers on the segment showing
the  Device  Address  and  Variable  Identity  for  each
one  -  see  Fig  5.      To  aid  identification,  operating
the  P  button  will  preview  the  value  of  the  data  at
each  selected  address,  if  required  it  may  then  be
allocated  to  one  of  the  eight  BA444DF-F  listener
inputs in-1, to in-8 by operating the P button again.
If  the  format  of  the  data  at  the  selected  Device
Address and Variable Identity is not DS-65 (float +
status) ‘no dAt’ will be displayed.

Fig 5  Device Address and Variable Identity

If  the  Device  Address  and  Variable  Identity  of  the
process variable to be displayed by the BA444DF-
F  indicator  are  known,  they  may  be  directly
entered  via  the  edit  sub-menu.    This  sub-menu
also  provides  full  control  of  both  the  numeric  and
bargraph display parameters.

Содержание FOUNDATION BA444DF-F

Страница 1: ...Issue 3 23 rd June 2011 BA444DF F FOUNDATION fieldbus Intrinsically safe Field mounting 8 variable Fieldbus Indicator Issue 3...

Страница 2: ...nter address display parameters 6 3 6 Enable selected input 6 3 7 Enter Device Address 6 3 8 Enter Variable Identity 6 3 9 Decimal point position 6 3 10 Configuration of bargraph 6 3 11 Scale numeric...

Страница 3: ...n alternative version of the fieldbus display order code BA444DF P for use on PROFIBUS PA networks 1 1 Documentation This instruction manual describes ATEX system design and installation of the BA444D...

Страница 4: ...ed comply with the Fieldbus Intrinsically Safe Concept FISCO defined in IEC 60079 Part 27 and also have separate entity parameters as shown below FISCO Entity Ui 17 5V dc 22 0V dc Ii 380mA dc 250mA dc...

Страница 5: ...any intrinsically safe segment providing The device powering the fieldbus segment is ATEX certified and has output parameters equal to or less than Uo 22V dc Io 250mA dc Po 1 2W The segment can provid...

Страница 6: ...F F enclosure is supplied with a bonding plate to ensure electrical continuity between the three conduit cable entries 5 2 Installation Procedure Fig 2 illustrates the instrument installation procedur...

Страница 7: ...nditioned as a fieldbus listener see section 6 3 13 to change it to a fieldbus node 6 2 1 Reset the indicator to node factory defaults rESEt The BA444DF F may be reset to the factory defaults that are...

Страница 8: ...the P button causes the BA444DF F indicator to search the fieldbus segment to which it is connected for all publishers Fig 5 shows how the indicator displays the hexadecimal Device Address and Variabl...

Страница 9: ...vice Address and Variable Identity Parameters can not be changed from within this function See section 6 3 4 Edit Enter addresses and display parameters Contains six sub functions that enable known fi...

Страница 10: ...10...

Страница 11: ...isplayed pressing E will allocate the selected process variable to the selected input and return the display to the Device Address and Variable Identity from which another process variable may be sele...

Страница 12: ...function pressing P again will reveal the existing bargraph justification The or button will scroll through the four options when the required justification is displayed pressing E will enter the sele...

Страница 13: ...ration menu using the or button To enter the function press P which will show that the instrument is configured as a fieldbus listener to change to a fieldbus node press the or button which will toggl...

Страница 14: ...normal operation after a few seconds the model number followed by the firmware version will be displayed prior to entering the display mode 7 2 Fault finding after commissioning ENSURE PLANT SAFETY BE...

Страница 15: ...After adding the required legends or fitting a new pre printed self adhesive escutcheon ensure that the gasket is correctly positioned before reassembling the instrument enclosure 8 2 Tag plate The BA...

Страница 16: ...e atmosphere in the form of a cloud of combustible dust in air is likely to occur occasionally in normal operation Zone 22 explosive atmosphere in the form of a cloud of combustible dust in air is not...

Страница 17: ...sion 1 or 2 Class I Group A B Group C Group D Class II Group E F G Class III Zone 0 1 or 2 Class 1 Group IIC Group IIB Group IIA The FM and cFM entity parameters are identical to the ATEX parameters a...

Страница 18: ...18...

Страница 19: ...19...

Страница 20: ...20...

Страница 21: ...21...

Страница 22: ...22...

Страница 23: ...23...

Страница 24: ...24...

Страница 25: ...fety parameters and installation requirements The ATEX system design requirements described in section 4 of this manual may therefore be used for IECEx installations in gas atmospheres but the local c...

Страница 26: ...ay be downloaded from www beka co uk www iecex com or requested from the BEKA sales office A3 3 Versions of the BA444DF F All versions of the BA444DF F fieldbus indicator have IECEx certification This...

Отзывы: