background image

18

8.  ACCESSORIES

8.1  Scale card

All models have a window on the right hand side of
the display through which to view a scale card
showing the units of measurement such as 

o

C,

mBar, RPM.   New indicators are fitted with a scale
card showing the units of measurement specified
when the indicator was ordered, if the units are not
specified a blank scale card will be fitted.   A pack
of scale cards pre-printed with common units of
measurement is available as an accessory.  These
can easily be fitted on-site to the indicator without
opening the indicator enclosure or removing it from
the panel,  See section 4.3 of this instruction
manual.

Custom scale cards for applications requiring less
common units of measurement are also available.

8.2  Tag information

New indicators are supplied with tag or application
information thermally printed onto the rear panel
adjacent to the terminals, legend as specified
when the indicator was ordered.  This tag
information is not visible from the front of the
instrument after installation.

8.3  Alarms

CAUTION

These alarms outputs should not be used for
critical safety applications such as an
emergency shut down system.

All models can be supplied with factory fitted dual
solid state, single pole alarm outputs.   Each alarm
output may be independently conditioned as a high
or low alarm with a normally open or normally
closed output in the non-alarm condition.

When the 4/20mA current powering the indicator is
removed both alarm outputs will open irrespective
of configuration.  The open circuit condition should
therefore be chosen as the alarm condition when
designing an alarm system.    Fig 10 illustrates the
conditions available and shows which are fail safe.

When an alarm occurs an alarm annunciator on
the indicator front panel is activated and if required
the numerical display can alternate between the
measured value and the alarm channel
identification ‘ALr1’ or ‘ALr2’.

CAUTION

The alarms are activated by the indicator’s
numerical display.  Use of the Tare Function
‘tArE’ will change the numerical display, the
alarms will continue to function at the original

displayed value, but this will correspond to a
different input current.

Fig 10  Alarm outputs

Configurable functions for each alarm include
adjustable setpoint, hysteresis, alarm delay and
alarm accept.

8.3.1  Solid state output

Each alarm has a galvanically isolated single pole
solid state switch output which as shown in Fig 11.
The output is polarised and current will only flow in
one direction.

Ron

=

less than 5

 + 0.7V

Roff

=

greater than 1M

Fig 11 Equivalent circuit of each alarm output

The solid state output of each alarm may be used
to switch any dc circuit with parameters equal to or
less than:

V

=

40V

I

=

200mA

Содержание BA507E

Страница 1: ...Supplied by Call us on 44 0 118 916 9420 Email info 247able com com Installation Maintenance Instructions BA507E BA508E BA527E BA528E...

Страница 2: ...rk Danehill Lower Earley Reading Berkshire RG6 4UT UK Tel 44 0 118 9311188 Email info able co uk Aberdeen Office Unit 6 Airside Business Park Kirkhill Industrial Estate Dyce Aberdeen AB21 0GT UK Tel 4...

Страница 3: ...Issue 3 1st December 2012 BA507E BA527E BA508E BA528E General Purpose loop powered panel mounting indicators issue 3...

Страница 4: ...error message 6 4 Lineariser under and over range 6 5 Lineariser default configuration 7 Maintenance 7 1 Fault finding during commissioning 7 2 Fault finding after commissioning 7 3 Servicing 7 4 Rout...

Страница 5: ...splay any variable represented by the 4 20mA current e g temperature flow pressure or level The BA507E BA527E BA508E and the BA528E are not certified for use in explosive atmospheres For hazardous are...

Страница 6: ...in display mode function has been enabled P E Access to configuration menu via optional security code Note BA527E and BA528E only If the indicator has been calibrated using the CAL function calibratio...

Страница 7: ...el providing that the operating temperature is between 40 C and 70 C Figs 4A and 4B show the overall dimensions of the 96 x 48mm and the 144 x 72mm instruments together with the recommended panel cut...

Страница 8: ...dovetail as shown in Fig 5 Push the knurled screw slightly forward to engage the thread and tighten by turning clockwise until it is just finger tight When both clamps are fitted ensure that the gaske...

Страница 9: ...rd unclip the protruding end of the flexible strip by gently pushing it upwards and pulling it out of the enclosure Peel the existing scale card from the flexible strip and replace it with a new print...

Страница 10: ...00 0000 Function FunC Std Std Display at 4mA Zero 0 0 0 00 Display at 20mA SPAn 100 0 100 00 Resolution rESn 1 digit 1 digit Bargraph start BarLo 0 00 Bargraph finish BarHi 100 00 P button in display...

Страница 11: ...value Using ZEro function set required display at 4mA by pressing pressing and to move to the next digit Similarly using SPAn function set required display at 20mA E P P E P E P E E Select type of ba...

Страница 12: ...ble lineariser To reveal the existing indicator function select FunC from the configuration menu and press P If the function is set as required press E to return to the menu or press the or button to...

Страница 13: ...ese conditions are not complied with the indicator displays FaiL and aborts the calibration c If the zero current is greater than the span current the instrument will be reverse acting i e an increasi...

Страница 14: ...between on and oFF When set as required press E to return to the tARE prompt in the configuration menu 5 10 Security code CodE Access to the instrument configuration menu may be protected by a four di...

Страница 15: ...lling the tank with known incremental volumes and calibrating the indicator to display the sum of the increments at each break point The number of break points required should first be entered using t...

Страница 16: ...14...

Страница 17: ...a new break point before the displayed break point The calibration of existing break points is not changed but the identification number of all subsequent break points is increased by one dEL Remove a...

Страница 18: ...ired press E to return to the 0 n prompt from which the the next break point can be selected using the or buttons When the required break point has been selected press P and set the required indicator...

Страница 19: ...s does not reveal the cause of the fault it is recommended that the instrument is replaced This can be done without disconnecting power but while the indicator is disconnected the 4 20mA loop will be...

Страница 20: ...output may be independently conditioned as a high or low alarm with a normally open or normally closed output in the non alarm condition When the 4 20mA current powering the indicator is removed both...

Страница 21: ...19...

Страница 22: ...section 8 3 3 SP1 Alarm setpoint 1 Adjusts the alarm setpoint The alarm is activated when the indicator display equals the setpoint See section 8 3 4 Hi Lo Alarm function Defines the alarm function a...

Страница 23: ...or fail safe operation both alarm outputs should be conditioned to be open in the alarm condition nC To check or change the alarm output status select no nC from the alarm configuration menu and press...

Страница 24: ...ving access to the indicator configuration menu Protection against accidental adjustment of the setpoints when the indicator is in the display mode is provided by a separate security code This direct...

Страница 25: ...plus the displayed value to be represented or a low and a high setpoint plus the displayed value to be represented by the bargraph as shown in Fig 15 Fig 15 Displayed value and setpoints on bargraph T...

Страница 26: ...backlight 8 4 1 Loop powering the backlight The backlight is loop powered by connecting it in series with the indicator 4 20mA input as shown in Fig 17 which increases the maximum indicator voltage dr...

Отзывы: