background image

1.  DESCRIPTION

The

BA414DF-F

Fieldbus

Indicator

is

an

intrinsically

safe,

FOUNDATION™

fieldbus

instrument compliant with ITK 6.3 that can display
one fieldbus process variable on a five digit LCD
and 31 segment analogue bargraph.

The

instrument is bus powered so no additional power
supply is required. 

            Communication             Fieldbus Function
                 Protocol                             Block

FOUNDATION™ fieldbus  Input Selector (1 x IS)

The Device Description files may be downloaded
from The FieldComm or the BEKA associates web
sites.

Housed in a robust IP66 glass reinforced polyester
(GRP) enclosure with a toughened glass window,
the BA414DF-F is surface mounting, or may be
pipe mounted using one of the accessory kits.  

The instrument has been certified intrinsically safe
by European Notified Body Intertek Testing and
Certification Ltd (ITS) for use in explosive gas and
combustible dust atmospheres.

ATEX dust

certification is an option – see Appendix 1.

For use in the USA and Canada, the BA414DF-F is
available

with

optional

intrinsic

safety

and

nonincendive FM and cFM Approval – see
Appendix 2. 

For international applications, all versions of the
BA414DF-F fieldbus indicator have IECEx intrinsic
safety approval allowing installation in explosive
gas atmospheres.

IECEx dust certification is

available as an option – see Appendix 3.

The instrument’s communication protocol is shown
on a label inside the terminal cover. The ‘-F’ order
code suffix also indicates the protocol but is not
shown on the instrument certification label.

1.1

  

Documentation

This instruction manual describes ATEX system
design and installation of the BA414DF-F Fieldbus
Indicator.

For commissioning information please

refer to:

FOUNDATION™ fieldbus

Fieldbus Interface Guide

for

Fieldbus Displays and

Fieldbus Indicators

 
which can be downloaded from the BEKA website
www.beka.co.uk

System design information for non-ATEX and dust
approvals is shown in appendices to this manual.

2.  INTRINSIC SAFETY CERTIFICATION

2.1 ATEX certificate

The BA414DF-F has been issued with an EC-Type
Examination Certificate ITS06ATEX25313X by
Notified Body Intertek Testing and Certification Ltd
(ITS) confirming compliance with harmonised
European standards.

The BA414DF-F fieldbus

indicator has Ex ia FISCO and entity parameter
certification plus Ex ic entity parameter approval
for use in Zone 2 with higher supply voltages.  

The EC-Type Examination certificate has been
used to confirm compliance with the European
ATEX Directive 2014/34/EU.

The BA414DF-F

carries the Community Mark and, subject to local
codes of practice, may be installed in any of the
European

Economic

Area

(EEA)

member

countries. ATEX certificates are also acceptable
for installations in Switzerland. 

This manual describes ATEX installations in
explosive gas atmospheres which conform with
EN

 

60079:14

Electrical

installation

design,

selection and erection.

When designing systems

for installation outside the UK, the local Code of
Practice should be consulted.  

For use in the presence of combustible dust,
please refer to Appendix 1.

2.2  Ex ia Zones, gas groups and T rating

The BA414DF-F has Group II Category 1G Ex ia
IIC T4 Ga –40 

 Ta 

 70

o

C FISCO and entity

parameter approval.

When connected to a

suitable certified system the BA414DF-F may be
installed in:

Zone 0

explosive gas air mixture
continuously present.

Zone 1

explosive gas air mixture 
likely to occur in normal 
operation.

Zone 2 explosive gas air mixture not

likely to occur, and if it does will
only exist for a short time.

Be used with gases in groups:

Group   A    propane
Group   B    ethylene
Group   C    hydrogen

In gases which may be used with equipment
having a temperature classification of:

T1

450ºC

T2

300ºC

T3

200ºC

T4

135ºC

At an ambient temperature between –40 and 
+70ºC  

3

Содержание BA414DF-F

Страница 1: ...Issue 9 27th March 2019 BA414DF F FOUNDATION fieldbus Intrinsically safe Field mounting Fieldbus Indicator Issue 9 ...

Страница 2: ...e 4 3 EMC 5 Maintenance 5 1 Fault finding during commissioning 5 2 Fault finding after commissioning 5 3 Servicing 5 4 Routine maintenance 5 5 Guarantee 5 6 Customer comments 6 Accessories 6 1 Scale and Tag marking 6 2 Tag plate 6 3 Pipe mounting kits 6 4 Fieldbus Interface Guide Appendix 1 ATEX dust certification Appendix 2 FM Approval for use in the USA and cFM Approval for use in Canada Appendi...

Страница 3: ... can be downloaded from the BEKA website www beka co uk System design information for non ATEX and dust approvals is shown in appendices to this manual 2 INTRINSIC SAFETY CERTIFICATION 2 1 ATEX certificate The BA414DF F has been issued with an EC Type Examination Certificate ITS06ATEX25313X by Notified Body Intertek Testing and Certification Ltd ITS confirming compliance with harmonised European s...

Страница 4: ...ut parameters as shown below FISCO Ex ia entity Ex ic entity Ui 17 5V 22 0V 32V Ii 380mA 250mA 125mA Pi 5 32W 1 2W 1W The maximum equivalent capacitance and inductance at terminals 1 2 is Ci 0nF Li 8µH 2 5 Certification Label Information The certification information label is fitted in a recess on the top outer surface of the enclosure It shows details of the ATEX certification a statement that th...

Страница 5: ...alent inductance is insignificant Therefore these BA414DF F parameters do not need to be considered 3 3 Ex ic entity systems The BA414DF F Fieldbus Indicator also has Ex ic certification with entity parameters for applications in Zone 2 The higher Ui voltage allows the indicator to be used with segment couplers powered by Ex e or Power i fieldbus trunks When installed in Zone 2 the BA414DF F Field...

Страница 6: ...ieldbus Indicator is 20 to 70o C 4 2 Installation Procedure Fig 2 illustrates the instrument installation procedure a Remove the instrument terminal cover by unscrewing the two captive A screws b Mount the instrument on a flat surface and secure with two M6 screws through the B holes Alternatively use one of the mounting kits described in section 6 3 c Remove the temporary hole plug and install a ...

Страница 7: ...es Status of fieldbus variable has a quality of BAD or a fault state is active Display has not yet received data Variable source Fieldbus configuration Bargraph scale flashes Variable is outside the limits defined for the bargraph Bargraph configuration All display segments activated Display is initialising This is normal operation after a few seconds the firmware version will be displayed prior t...

Страница 8: ...display escutcheon remove the terminal cover by unscrewing the two A screws which will reveal two concealed D screws Unscrew all four D screws and carefully lift off the front of the instrument The location of all these screws is shown in Fig 2 After adding the required legends or fitting a new pre printed self adhesive escutcheon ensure that the gasket is correctly positioned before reassembling ...

Страница 9: ...ick Dust layer on BA414DF F Refer to over 5mm thick EN 60079 14 At an ambient temperature between 20 and 60o C The Indicator may be connected to any intrinsically safe segment providing The device powering the fieldbus segment is ATEX Ex ia IIIC certified for Zone 20 21 or 22 applications or Ex ib IIIC certified for application in Zone 21 or 22 The output parameters should be equal to or less than...

Страница 10: ...FISCO Field Device plus BEKA associates name and location IECEx dust approval information is also shown non European certification information may also be included The label includes boxed areas which should be marked by the installer to show which of the certifications is being used The instrument serial number and date of manufacture are recorded on a separate label inside the terminal compartme...

Страница 11: ... 1 or 2 Class I Group A B Group C Group D Class II Group E F G Class III Zone 0 1 or 2 Class 1 Group IIC Group IIB Group IIA The FM and CFM entity parameters are identical to the ATEX parameters and like the ATEX certification confirm that the BA414DF F complies with the FISCO Field Device requirements specified in IEC60079 27 The intrinsically safe system shown in Fig 1 of this manual may therefo...

Страница 12: ...12 ...

Страница 13: ...13 ...

Страница 14: ...14 ...

Страница 15: ...15 ...

Страница 16: ...16 ...

Страница 17: ...17 ...

Страница 18: ...18 ...

Страница 19: ...llation requirements for both gas and dust approvals The ATEX installation requirements specified in the main section and in Appendix 1 of this manual may therefore be used for IECEx installations but the local code of practice should also be consulted CAUTION installation in Zone 0 When installed in a Zone 0 potentially explosive atmosphere requiring EPL Ga apparatus the instrument shall be insta...

Отзывы: