A Quick Guide to the Use of this Instruction Manual

Your BA374NC has been factory set to the requirements specified on your purchase order.   If
no requirements were given, the instrument will be supplied with default settings i.e.  3 wire
RTD input; high resolution display in °C; 4 to 20mA output corresponding to an input and
display of 0.0 to 100.0°C.

If the BA374NC is set to your requirements before putting it into use you should refer to: 
                                          
                                         Section 4               Electrical System Design

                                         Section 5               Installation

                                         Section 2.1            Controls

If optional alarms are fitted you should refer to Section 11. You will also find the 'Trim' menu
described in Section 7 useful if you wish to calibrate the complete control loop.

If the BA374NC has been supplied with default settings and you wish to change these you will
need to refer to:

                                         Section 6               Programming and Calibration

This section explains the programme functions of the push-buttons and leads you through the
various stages of programming the unit. Among the sub-divisions of Section 6 you will find
information on setting up for

                                         Section 6.3            Thermocouple input

                                         Section 6.4            Resistance thermometer input 

                                         Section 6.5            Voltage input

A word of warning:

 Section 6.8 explains how you can calibrate the instrument's internal

references. It is only necessary to carry out this operation if you will be using these internal
references to calibrate the instrument display, 

i.e. not using an external calibrator,

 and then

only annual or less frequent adjustment is required.

 It is not necessary to calibrate these

references before putting the unit into operation, 

we've already done that for you. For this

reason the calibration section of the programme menu is protected by an additional security
code. You should only enter this section of the programme if you are sure that you want to
carry out this calibration.

By following the notes in this manual you should be able to set up your BA374NC without any
problem. If you do run into difficulties however Section 9 (Maintenance) provides some
guidance on typical problems you may find during commissioning.  If you still have difficulties
please contact our Sales Department

 

(01462 438301). Our Sales Engineers will be able to talk

you through the procedure.

Содержание BA374NC

Страница 1: ...BA374NC Type nL certified field mounting indicating temperature transmitter Issue 5 3rd December 2003...

Страница 2: ...brating the internal references 6 8 1 Conditioning internal references 6 8 2 Voltage input reference 6 8 3 3 wire RTD reference 6 8 4 4 wire RTD reference 6 8 5 Output current reference 7 Trim Menu 7...

Страница 3: ...tion 6 you will find information on setting up for Section 6 3 Thermocouple input Section 6 4 Resistance thermometer input Section 6 5 Voltage input A word of warning Section 6 8 explains how you can...

Страница 4: ...4...

Страница 5: ...mperature measurement applies linearisation using look up tables stored in permanent memory The resulting temperature is shown on the internal display The 4 20mA output current is proportional to the...

Страница 6: ...components which may generate a spark in normal operation e g switches and switches and potentiometers These approvals confirm that the BA374NC is safe in normal operation and may be installed in Zone...

Страница 7: ...res are surface mounting but may be pipe or panel mounted using the accessory kits described in section 10 Fig 3 BA374NC installation procedure To simplify installation the enclosure may be installed...

Страница 8: ...accuracies Each function in the Programme Menu is summa rised in section 6 1 and includes a cross reference to more detailed information in the following sections Although this simple menu driven syst...

Страница 9: ...9...

Страница 10: ...10...

Страница 11: ...can be reached by scrolling through the functions using the Up and Down push buttons as shown in Figs 5A and 5B When returning to the display mode following changes to any parameters the transmitter w...

Страница 12: ...See section 6 2 1 For Thermocouple Inputs tYPE Sensor type Selects the type of thermocouple input See section 6 3 1 dEg Units of display Sets display to o C or o F See section 6 3 2 rESn Display reso...

Страница 13: ...pressing P will reveal a sub menu from which for thermocouple and resistance thermometer inputs the type of sensor may be selected and associated parameters programmed 6 3 For thermocouple inputs 6 3...

Страница 14: ...eit To check or change the display units select dEg from the sub menu and press P to reveal the current setting The setting may be changed by pressing the Up or Down button When the required unit is d...

Страница 15: ...20mA output current The transmitter output current is derived from the transmitter display and will change if the display calibration is changed ZEro defines the display at which the output is 4mA an...

Страница 16: ...ined in a protected sub menu to minimise the possibility of accidental adjustment Select Cond from the menu and press P The transmitter will display 0000 Using the Up and Down buttons and the P button...

Страница 17: ...Pressing P will change the display to 020A and the transmitter output current to approximately 20mA The transmitter output current shown on the external current measuring instrument should be adjusted...

Страница 18: ...re of the primary element operate the Up or Down buttons until the transmitter display is correct Immediately the Up or Down button is operated the annunciator in the top left hand corner of the displ...

Страница 19: ...uple With tHC displayed press P to enter the thermocouple sub menu The first parameter tYPE will be displayed pressing P again will give access to the seven different types of thermocouple Select J us...

Страница 20: ...be less accurate With the transmitter displaying the ZEro prompt and the correct input connected to the transmitter push P which will cause the transmitter to display the thermocouple temperature Adju...

Страница 21: ...osure Before replacing the instrument front cover ensure that the sealing gasket is undamaged and is free from dirt and foreign bodies If a BA374NC fails after it has been functioning correctly the pr...

Страница 22: ...eel and the kit contains a trim plate to conceal the edge of the panel aperture 11 0 ALARMS The BA374NC can be supplied with two solid state single pole alarm outputs Each output may be programmed as...

Страница 23: ...gain only the functions on alarm 1 are listed but alarm 2 has identical facilities 11 2 Summary of programmable alarm functions Display Summary of function EnbL Alarm enable Enables or disables the al...

Страница 24: ...24 Fig 11 Programme structure for optional alarms...

Страница 25: ...rm output status select no nc from the alarm sub menu and press P reveal the current setting The function can be changed by pressing the Up or Down button followed by the E button to return the no nc...

Страница 26: ...cSP prompt in the Programme Menu Code 0000 will disable the security code allowing direct access to the setpoints by pressing the P and Up buttons simultaneously New instruments are supplied with this...

Страница 27: ...input safety parameters are Ui 30V dc Ii 100mA Pi 1 3W This allows the backlight to be powered from a safe area 18 to 30V dc supply as shown in Fig 13 Providing the supply voltage in normal operation...

Отзывы: