Step 10 Return to the display mode

With the 'CAL' prompt displayed, press

E

 to store the conditioning and

calibration information and return to the
operating mode.  All the display
decimal points will be activated while
the information is being stored. This
completes the calibration of the
transmitter using the internal
references.

The following steps explain how, using the Trim
Menu, the BA374NC transmitter may be calibrated
against an external reference, such as a voltage
source or temperature calibrator. Alternatively, the
complete loop including the primary element may
be calibrated by cooling and heating the
thermocouple to -10 and +240

o

C.

Step 11 Entering the Trim Menu

Enter the Trim Menu by simultaneously
pressing the

 P

 and 

Down

 buttons - see

Fig 7. The transmitter will respond by
displaying 'COdE' .  Press 

P

 to clear

this prompt and set the display to the
security code 1209 using the 

Up

Down

and 

P

 push-buttons.  Pressing 

E

 will

enter the code.  After a few seconds
during which all the decimal points will
be activated, the first parameter 'ZEro'
will be displayed.

Step 12 Zero adjustment

This adjustment trims the display to
read exactly -10.0

o

C and the output

current to be exactly 4.000mA when the
type J  thermocouple is at  -10

o

C.  This

can be achieved by cooling the
thermocouple to  -10

o

C or by inputting

the voltage which the thermocouple
should theoretically produce.  If a
voltage is used the transmitter cold
junction compensation (see section
6.3.4) should be turned off during the
calibration, or the input voltage should
be corrected for ambient temperature,
although this may be less accurate.

With the transmitter displaying the
'ZEro' prompt and the correct input
connected to the transmitter, push 

P

which will cause the transmitter to
display the thermocouple temperature.
Adjust the display with the 

Up

 and

Down

 buttons until it is -10.0 

Note:  Immediately the 

Up

 or 

Down

button is operated the annunciator in
the top left hand corner of the display
will be activated to warn that any
change in the input signal will be
ignored while this display correction is
being made.  

Operating the  

 push-button will lock

the transmitter display at -10.0

 

and

allow the output current to be trimmed
using the 

Up

 or 

Down

 button.  When

the output current measured by a DVM
is 4.000mA, the 

E

 push-button should

be operated to return the transmitter to
the 'ZEro' prompt.
This completes calibration at -10.0

o

C.

Step 13 Span adjustment

With the transmitter displaying the
'ZEro' prompt select 'SPAn' by pressing
the 

Up

 or 

Down

 button.  With the

thermocouple heated to 240

o

C, or an

equivalent voltage input, push 

P

 which

will cause the transmitter to display the
thermocouple temperature. Adjust the
display with the 

Up

 and 

Down

 buttons

until it is 240.0   

Note:  Immediately the 

Up

 or 

Down

button is operated the annunciator in
the top left hand corner of the display
will be activated to warn that any
change in the input signal will be
ignored while this display correction is
being made.  

Operating the  

 push-button will lock

the transmitter display at 240.0

 

and

allow the output current to be trimmed
using the 

Up

 or 

Down

 button.  When

the output current measured by a DVM
is 20.000mA, the 

E

 push-button should

be operated to return the transmitter to
the 'SPAn' prompt.
This completes calibration at 240.0

o

C.

Pressing the 

push-button again will

return the transmitter to the display
mode.

      20

Содержание BA374NC

Страница 1: ...BA374NC Type nL certified field mounting indicating temperature transmitter Issue 5 3rd December 2003...

Страница 2: ...brating the internal references 6 8 1 Conditioning internal references 6 8 2 Voltage input reference 6 8 3 3 wire RTD reference 6 8 4 4 wire RTD reference 6 8 5 Output current reference 7 Trim Menu 7...

Страница 3: ...tion 6 you will find information on setting up for Section 6 3 Thermocouple input Section 6 4 Resistance thermometer input Section 6 5 Voltage input A word of warning Section 6 8 explains how you can...

Страница 4: ...4...

Страница 5: ...mperature measurement applies linearisation using look up tables stored in permanent memory The resulting temperature is shown on the internal display The 4 20mA output current is proportional to the...

Страница 6: ...components which may generate a spark in normal operation e g switches and switches and potentiometers These approvals confirm that the BA374NC is safe in normal operation and may be installed in Zone...

Страница 7: ...res are surface mounting but may be pipe or panel mounted using the accessory kits described in section 10 Fig 3 BA374NC installation procedure To simplify installation the enclosure may be installed...

Страница 8: ...accuracies Each function in the Programme Menu is summa rised in section 6 1 and includes a cross reference to more detailed information in the following sections Although this simple menu driven syst...

Страница 9: ...9...

Страница 10: ...10...

Страница 11: ...can be reached by scrolling through the functions using the Up and Down push buttons as shown in Figs 5A and 5B When returning to the display mode following changes to any parameters the transmitter w...

Страница 12: ...See section 6 2 1 For Thermocouple Inputs tYPE Sensor type Selects the type of thermocouple input See section 6 3 1 dEg Units of display Sets display to o C or o F See section 6 3 2 rESn Display reso...

Страница 13: ...pressing P will reveal a sub menu from which for thermocouple and resistance thermometer inputs the type of sensor may be selected and associated parameters programmed 6 3 For thermocouple inputs 6 3...

Страница 14: ...eit To check or change the display units select dEg from the sub menu and press P to reveal the current setting The setting may be changed by pressing the Up or Down button When the required unit is d...

Страница 15: ...20mA output current The transmitter output current is derived from the transmitter display and will change if the display calibration is changed ZEro defines the display at which the output is 4mA an...

Страница 16: ...ined in a protected sub menu to minimise the possibility of accidental adjustment Select Cond from the menu and press P The transmitter will display 0000 Using the Up and Down buttons and the P button...

Страница 17: ...Pressing P will change the display to 020A and the transmitter output current to approximately 20mA The transmitter output current shown on the external current measuring instrument should be adjusted...

Страница 18: ...re of the primary element operate the Up or Down buttons until the transmitter display is correct Immediately the Up or Down button is operated the annunciator in the top left hand corner of the displ...

Страница 19: ...uple With tHC displayed press P to enter the thermocouple sub menu The first parameter tYPE will be displayed pressing P again will give access to the seven different types of thermocouple Select J us...

Страница 20: ...be less accurate With the transmitter displaying the ZEro prompt and the correct input connected to the transmitter push P which will cause the transmitter to display the thermocouple temperature Adju...

Страница 21: ...osure Before replacing the instrument front cover ensure that the sealing gasket is undamaged and is free from dirt and foreign bodies If a BA374NC fails after it has been functioning correctly the pr...

Страница 22: ...eel and the kit contains a trim plate to conceal the edge of the panel aperture 11 0 ALARMS The BA374NC can be supplied with two solid state single pole alarm outputs Each output may be programmed as...

Страница 23: ...gain only the functions on alarm 1 are listed but alarm 2 has identical facilities 11 2 Summary of programmable alarm functions Display Summary of function EnbL Alarm enable Enables or disables the al...

Страница 24: ...24 Fig 11 Programme structure for optional alarms...

Страница 25: ...rm output status select no nc from the alarm sub menu and press P reveal the current setting The function can be changed by pressing the Up or Down button followed by the E button to return the no nc...

Страница 26: ...cSP prompt in the Programme Menu Code 0000 will disable the security code allowing direct access to the setpoints by pressing the P and Up buttons simultaneously New instruments are supplied with this...

Страница 27: ...input safety parameters are Ui 30V dc Ii 100mA Pi 1 3W This allows the backlight to be powered from a safe area 18 to 30V dc supply as shown in Fig 13 Providing the supply voltage in normal operation...

Отзывы: