background image

22

9.4  Lineariser

The indicator can be supplied with a sixteen point
lineariser which may be adjusted to compensate
for almost any non linear variable.  For example, a
level signal from a horizontal cylindrical tank may
be linearised by the indicator to display the tank
contents in linear volumetric units.

The addition of linearising software does not affect
the intrinsic safety of the indicator.

Fig 13 shows a typical linearising characteristic.
Up to sixteen break-points may be programmed to
occur at any input current between 4 and 20mA.
The slope between adjacent break-points may be
set anywhere between -1250 and +1250 display
counts per milliamp.  Greater slopes may be
programmed, but the indicator performance will be
degraded.  A linear characteristic can be obtained
by programming just two points, one at 4mA and
the other at 20mA.

Fig 13   Typical indicator characteristic

The lineariser software does not change the main
programme menu, but the CAL and SEt functions
are extended as shown in  Fig 14.    As with a
linear indicator, calibration may be performed with
an external calibrator using the 'CAL' function,  or
from the internal references using the 'Set'
function.

9.4.1  Calibration using an external source

This method allows direct calibration with a current
source, and  is preferred  when traceability is
required.   If the exact system non-linearity is
unknown, the method also allows direct calibration
from the variable to be displayed.  e.g. the output
from a level sensor in an irregular tank may be
displayed in linear volumetric units by filling the
tank with known incremental volumes and
calibrating the indicator to display the sum of the
increments at each break-point.

The number of break-points required should first
be entered using the Add and dEL  functions.  In
both these functions the indicator initially displays
the current break-point and the total number of
break-points being used. e.g.

              

     5               13

                   current                total number  of
                break-point               break-points

'Add'

Add a break-point

Adds a new break-point before the
displayed break-point.  The calibration
of existing break-points is not changed,
but the identification number of all
subsequent break-points is increased
by one.

'dEL'

'Remove a break-point

Removes the displayed break-point
and joins the preceding break-point to
the following break-point with a straight
line.   The identification number of all
subsequent break-points is decreased
by one.

To add a break-point select 'CAL' from the main
menu and press P  to enter the 'Add' function;
press  P  again  to reveal the current and total
number of break-points.    Each subsequent
operation of the  P  push-button will introduce an
additional break-point.  When adding a break-point
to a calibrated indicator, the insertion position can
be selected using the Up and Down push-buttons.

The delete break-point function dEL operates in a
similar manner to the Add function described
above.

When the required number of break-points has
been entered, return to the sub-menu by pressing
E.  The indicator will display 'Add' or 'dEL'
depending upon the last function used.  Each
break-point can now be programmed

Select  'PtS' from the sub-menu and press P which
will select the first break-point '0  n', where n is the
total number of break-points entered.  The
selected break-point can be changed using the Up
and Down buttons.  When the required break-point
has been selected press P.  Set the indicator input
current to the exact value at which the break-point
is to occur, and adjust the indicator display using
the Up and Down buttons and P to move between
digits.  When the required display has been set,
press  E  to enter the information and return to the
sub-menu from which another break-point  can be
selected.

Содержание BA327C

Страница 1: ...BA327C BA328C intrinsically safe loop powered 4 digit panel mounting indicators issue 11 Issue 11 3rd November 2009...

Страница 2: ...rator 8 Maintenance 8 1 Fault finding during commissioning 8 2 Fault finding after commissioning 8 3 Servicing 8 4 Routine maintenance 8 5 Guarantee 8 6 Customer comments 9 Accessories 9 1 Scale card...

Страница 3: ...lt conditions the voltage current and power at the input terminals of each instrument will not exceed those specified for simple apparatus in Clause 5 4 of EN50020 1994 The BA327C and BA328C are also...

Страница 4: ...allations in Switzerland This manuals describes installations which conform with EN60079 Part 14 When designing systems for installation outside the UK th local Code of Practice should be consulted 3...

Страница 5: ...s BEKA associates name and location Non European certification information may also be shown 4 SYSTEM DESIGN FOR HAZARDOUS AREAS 4 1 Transmitter loops BA327C and BA328C indicators may be connected in...

Страница 6: ...iers a two channel Zener barrier or a galvanic isolator must be used Again it is necessary to ensure that the voltage capability of the 4 20mA signal is sufficient to drive the indicator plus the volt...

Страница 7: ...erall dimensions of each instrument and the required panel cut out dimensions To achieve an IP65 seal between the instrument enclosure and the instrument panel the smaller tolerance aperture must be u...

Страница 8: ...3 EMC The BA327C and BA328C comply with the requirements of the European EMC Directive 2004 108 EEC For specified immunity all wiring should be in screened twisted pairs with the screens earthed with...

Страница 9: ...ating decimal points for a few seconds while the new information is stored in non volatile memory All new BA327C and BA328C indicators are supplied calibrated as requested at the time of ordering If c...

Страница 10: ...10...

Страница 11: ...lowing table shows the output current from a non linearised differential flowmeter of full flow Current output mA 2 5 4 01 10 0 4 16 25 0 5 00 50 0 8 00 75 0 13 00 100 0 20 00 6 3 Resolution rESn This...

Страница 12: ...orrect pressing P will transfer control to the next digit When the least significant digit has beenadjusted press E to return to the SPAn prompt followed by E to return to the menu 6 7 Conditioning su...

Страница 13: ...calibration and conditioning of the instrument may be protected by a four digit security code which must be entered before access to the programme menu is granted New instruments are programmed with...

Страница 14: ...Down buttons select 50 and then press E twice to return to the main menu Step 5 Define function of P push button Select C P from the main menu and press P to reveal the function of the P button in the...

Страница 15: ...to the main menu Step 4 Define function of P push button Select C P from the main menu and press P to reveal the function of the P button in the display mode Select 4 20 and return to the main menu by...

Страница 16: ...ing correctly the following procedure should be followed Symptom Cause Solution No display No power Check that current 0V between supply is flowing in the loop terminals 1 and 3 Unstable 4 20mA input...

Страница 17: ...ependently programmed as high or low alarms with normally open or normally closed outputs Fig 8 illustrates the conditions available and shows which are fail safe i e in the alarm condition the output...

Страница 18: ...18...

Страница 19: ...mmable alarm functions Display Description of function EnbL Alarm enable Enables or disables the alarm function without changing the alarm parameters See section 9 3 4 SP1 Alarm setpoint 1 Adjusts the...

Страница 20: ...y pressing the Up or Down button followed by the E button to return to the alarm menu 9 3 8 Hysteresis HStr During programming hysteresis is shown in the units the indicator has been calibrated to dis...

Страница 21: ...security code allowing direct access to the setpoints by pressing the P and Up buttons simultaneously New instruments with alarms are supplied with this function disabled and the security code set to...

Страница 22: ...using the Add and dEL functions In both these functions the indicator initially displays the current break point and the total number of break points being used e g 5 13 current total number of break...

Страница 23: ...f break points Each subsequent operation of the P push button will introduce an additional break point When adding a break point to a calibrated indicator the insertion position can be selected using...

Страница 24: ...24...

Страница 25: ...e output which enhances daylight viewing but an additional IS interface and field wiring are required 9 6 1 Separately powered backlight The separately powered backlight is segregated from the measuri...

Страница 26: ...is loop powered This eliminates the need for a separate Zener barrier or galvanic isolator and associated wiring for the backlight thus significantly reducing the installation cost As shown in Fig 16...

Страница 27: ...3 A1 2 4 20mA input in nonincendive circuit The BA327C and BA328C are also Factory Mutual approved as nonincendive for Class I Division 2 locations which allows both indicators to be installed in Div...

Страница 28: ...between the backlight and the measuring circuits should be maintained by following the wiring methods recommended in ANSI NFPA70 article 504 See Fig 15 A1 3 2 2 Loop powered backlight This backlight h...

Страница 29: ...29...

Страница 30: ...30...

Страница 31: ...31...

Страница 32: ...f Conformity number IECEx ITS 05 0003 which specifies the following certification code and marking Ex ia IIC T5 Ta 40o C to 60 o C The specified intrinsic safety parameters are identical to the ATEX s...

Отзывы: