BEKA BA307E-SS Скачать руководство пользователя страница 11

6.  CONFIGURATION AND CALIBRATION

Indicators are configured and calibrated via the four
front  panel  push-buttons.      All  the  configuration
functions  are  contained  in  an  easy  to  use  intuitive
menu that is shown diagrammatically in Fig 10.

Each menu function is summarised in section 6.1 and
includes  a  reference  to  more  detailed  information.
When  the  indicator  is  fitted  with  alarms  additional
functions are added to the menu which are described
in section 9.3

Throughout this manual push buttons are shown as 

P

,

E

,

   

 

or  

, and legends displayed by the indicator

are shown  within inverted  commas  e.g. 'CAL' and '
ALr2'.

Access  to  the  configuration  menu  is  obtained  by
operating the 

P

 and 

E

 push buttons simultaneously.  If

the indicator security code is set to the default 0000
the  first  parameter  'FunC'  will  be  displayed.    If  a
security code other than the default code 0000 has
already been entered, the indicator will display 'CodE'.
Pressing the  

P

 button will clear this prompt allowing

each digit of the code to be entered using the 

 and

 push buttons and the  

P

 button to move control to

the next digit.  When the correct four digit code has
been entered pressing

 

E

 will cause the first parameter

'FunC' to be displayed.   If the code is incorrect, or a
button  is  not  pressed  within  twenty  seconds,  the
indicator will automatically return to the display mode.

Once  within  the  configuration  menu  the  required
parameter  can  be  reached  by  scrolling  through  the
menu using the 

 and 

 

push-buttons as shown in

Fig 10.  When returning to the display mode following
recalibration or a change to any function, the indicator
will display  ‘dAtA’ followed by ‘SAVE’ while the new
information is stored in non-volatile memory.

All  new  indicators  are  supplied  calibrated  as
requested at the time of ordering.  If calibration is not
requested, indicators will be configured as follows:

BA307E-SS BA327E-SS

Access code  ‘CodE’

0000

0000

Function  ‘FunC’

Std

Std

Display at 4mA  ‘Zero’

0.0

0.00

Display at 20mA  ‘SPAn’

100.0

100.00

Resolution  ‘rESn’

1 digit

1 digit

Bargraph start  ‘BarLo’

-----

0.00

Bargraph finish  ‘BarHi’

-----

100.00

P

 button in display mode ‘C—P’

%

%

Tare  ‘tArE’

Off

Off

Default configuration can easily be changed on-site.

6.1  Summary of configuration functions

This  section  summarises  each  of  the  main
configuration  functions  and  includes  a  cross
reference to a more detailed description.  Fig 10
illustrates the location of each function within the
configuration  menu.    The  lineariser  and  the
optional  factory  fitted  alarms  are  described
separately in sections 7 and 9.3 of this manual.

  

Display

      Summary of function

'FunC'

Indicator function

Defines  the  relationship  between  the
4/20mA input current and the indicator
display.  May be set to:

‘Std’

Standard linear relationship

‘root’

Square root extraction

‘Lin’

16 segment adjustable 
lineariser – see section 7.

See section 6.2

'rESn'

Display resolution

Defines  the  resolution  of  the  least
significant display digit.  May be set to
‘1’, ‘2’, ‘5’ or ‘10’ digits.
See section 6.3

'dP'

Decimal point

Positions  a  dummy  decimal  point
between any of the digits or turns it off.
See section 6.4

'CAL'

Calibration  of  the  digital  display
using an external current source.

Enables  the  zero  and  span  of  the
indicator  to  be  adjusted  using  an
external  current  source  such  as  a
calibrator.  When used with an accurate
traceable  current  source  this  is  the
preferred method of calibration.
See section 6.5

'SEt'

Calibration of display using internal
references.

Enables  the  zero  and  span  of  the
indicator  to  be  adjusted  without  the
need  for  an  accurate  input  current  or
disconnection  from  the  4/20mA  loop.
See section 6.6

'bAr'

Bargraph format and calibration 

Only the BA327E -SS has a bargraph

The  bargraph  may  be  conditioned  to
start  from  left,  right  or  centre  of  the
display, or it may be disabled.  When
optional  alarms  are  fitted  it  can  also
display  both  alarm  setpoints  and  the
measured value. 
The bargraph may be calibrated to start
and  finish  at  any  value  within  the
indicator’s calibrated digital display.
See section 6.7

11

Содержание BA307E-SS

Страница 1: ...Issue 3 4th May 2017 BA307E SS BA327E SS intrinsically safe loop powered rugged panel mounting indicators Issue 3 ...

Страница 2: ...7 2 Lineariser calibration using internal reference 7 2 1 Example adding break points to a new indicator 7 3 Under and over range 7 4 Lineariser under and over range 7 5 Lineariser default configuration 8 Maintenance 8 1 Fault finding during commissioning 8 2 Fault finding after commissioning 8 3 Servicing 8 4 Routine maintenance 8 5 Guarantee 8 6 Customer comments 9 Accessories 9 1 Scale card 9 2...

Страница 3: ...nd confirmed that the front of both instruments comply with Ex e Ex p and Ex t ingress and impact requirements The intrinsic safety EC Type Examination certificate specifies that under fault conditions the output voltage current and power at the 4 20mA input terminals will not exceed those specified for simple apparatus in Clause 5 7 of EN 60079 11 which simplifies installation and documentation F...

Страница 4: ...x gas and dust certification This section of the instruction manual describes ATEX gas certification ATEX dust and IECEx approvals are described in Appendixes 1 and 2 3 1 ATEX gas certification Both models have been issued with common EC Type Examination Certificate ITS14ATEX28077X by Notified Body Intertek Testing and Certification Ltd This confirms compliance with harmonised European standards a...

Страница 5: ...e and inductance between the two 4 20mA input terminals 1 and 3 is Ci 13nF Li 8µH Effectively 0 The maximum permitted loop cable parameters can be calculated by adding these figures to Ci and Li of other instruments in the loop and subtracting the totals from the maximum cable capacitance Co and cable inductance Lo permitted for the Zener barrier or galvanic isolator powering the loop Although the...

Страница 6: ... 2 wire transmitter powered by a Zener barrier and alternatively by a galvanic isolator Fig 2a Loop powered by a Zener barrier Fig 2b Loop powered by a galvanic isolator 4 2 Remote indication The indicators may be driven via an intrinsically safe interface from a 4 20mA safe area signal to provide a remote display within a hazardous area The type of intrinsically safe interface is not critical eit...

Страница 7: ...and impact protection as the front of both indicators comply with Ex e impact and ingress requirements Although mounted in an Ex e panel enclosure the BA307E SS or BA327E SS remains Group II Category 1G Ex ia IIC T5 Ga intrinsically safe apparatus so must be powered via a Zener barrier or galvanic isolator as described in section 4 1 and 4 2 of this manual Some Zener barriers and galvanic isolator...

Страница 8: ...bed in section 4 1 and 4 2 of this manual to ensure that the instrument s front panel push button switches are nonincendive When installed in an Ex p panel enclosure the four vents at the rear of indicator which are shown in Fig 8 should not be obstructed Zener barriers and galvanic isolators may be installed in the same Ex p enclosure as the indicator All may be mounted in an Ex px enclosure inst...

Страница 9: ...the edges of aperture are de burred b Inspect the indicator s captive gasket and ensure that it is not damaged before inserting the indicator into the panel enclosure aperture c If the enclosure panel is less than 1 0mm thick or is non metallic an optional BEKA stainless steel support plate should be slid over the rear of the indicator before the panel clamps are fitted to evenly distribute the cl...

Страница 10: ...requested units of measurement if this information is not supplied when the indicator is ordered a blank card will be fitted A pack of self adhesive scale cards printed with common units of measurement is available as an accessory from BEKA associates Custom printed scale cards can also be supplied To change a scale card unclip the protruding end of the flexible strip by gently pushing it upwards ...

Страница 11: ...0 Function FunC Std Std Display at 4mA Zero 0 0 0 00 Display at 20mA SPAn 100 0 100 00 Resolution rESn 1 digit 1 digit Bargraph start BarLo 0 00 Bargraph finish BarHi 100 00 P button in display mode C P Tare tArE Off Off Default configuration can easily be changed on site 6 1 Summary of configuration functions This section summarises each of the main configuration functions and includes a cross re...

Страница 12: ...12 ...

Страница 13: ...e lineariser To reveal the existing indicator function select FunC from the configuration menu and press P If the function is set as required press E to return to the menu or press the or button to change the setting followed by the E button to return to the configuration menu Std Linear Provides a linear relationship between the 4 20mA indicator input current and the indicator display root Square...

Страница 14: ...mA If these conditions are not complied with the indicator displays FaiL and aborts the calibration c If the zero current is greater than the span current the instrument will be reverse acting i e an increasing input current will cause the display to decrease 6 6 Calibration using internal reference SEt Using the SEt function the indicator can be calibrated without the need for an accurate externa...

Страница 15: ...tting between on and oFF When set as required press E to return to the tARE prompt in the configuration menu 6 10 Security code CodE Access to the instrument configuration menu may be protected by a four digit security code which must be entered to gain access New instruments are configured with the default security code 0000 which allows unrestricted access to all configuration functions To enter...

Страница 16: ...irect calibration of the lineariser with an external current source and is the preferred method when traceability is required If the exact system non linearity is unknown this method also allows direct calibration from the variable to be displayed e g the output from a level sensor in an irregular tank may be displayed in linear volumetric units by filling the tank with known incremental volumes a...

Страница 17: ...17 ...

Страница 18: ...lowing break point 7 1 1 Example Adding break points to a new indicator When adding break points to a new indicator or to a lineariser following resetting to the factory defaults using the LtAb function described in section 6 11 each additional break point should be added before the highest existing breakpoint The first additional break point should be added before the default break point 1 1 whic...

Страница 19: ...he corresponding indicator display at each of the break points can now be defined using the diSP sub function Using the and buttons select the diSP sub function and press P which will reveal the starting point for the first break point 0 n where n is the total number of break points entered Press P and use the and buttons and the P button to move between digits to enter the required indicator disp...

Страница 20: ...el enclosure additional precautions may be required If an indicator fails after it has been functioning correctly follow the procedure shown in section 8 1 If this does not reveal the cause of the fault it is recommended that the instrument is replaced This can be done without disconnecting power but while the indicator is disconnected the 4 20mA loop will be open circuit 8 3 Servicing All BA307E ...

Страница 21: ...rent powering the indicator is removed both alarm outputs will open irrespective of configuration The open circuit condition should therefore be chosen as the alarm condition when designing an alarm system Fig 13 illustrates the conditions available and shows which are fail safe When an alarm occurs an alarm annunciator on the indicator front panel is activated and if required the numerical displa...

Страница 22: ...22 ...

Страница 23: ...l facilities Summary of alarm configuration functions Display Description of function EnbL Alarm enable Enables or disables the alarm without changing the alarm parameters See section 9 3 4 SP1 Alarm setpoint 1 Adjusts the alarm setpoint The alarm is activated when the indicator display equals the setpoint See section 9 3 5 Hi Lo Alarm function Defines the alarm function as High or Low See section...

Страница 24: ... fail safe operation both alarm outputs should be conditioned to be open in the alarm condition nC To check or change the alarm output status select no nC from the alarm configuration menu and press P to reveal the setting The function may be changed by pressing the or button followed by the E button to return to the alarm configuration menu 9 3 8 Hysteresis HStr Hysteresis is shown in the units t...

Страница 25: ...ng access to the indicator configuration menu Protection against accidental adjustment of the setpoints when the indicator is in the display mode is provided by a separate security code This direct setpoint access menu is enabled and the separate security code entered from the ACSP function in the alarm configuration menu as shown in Fig 15 To change the menu parameters select ACSP from the config...

Страница 26: ...layed value to be represented or a low and a high setpoint plus the displayed value to be represented by the bargraph as shown in Fig 18 Fig 18 Displayed value and setpoints on bargraph The bargraph area below the low alarm setpoint SP1 and the bargraph area above the high alarm setpoint SP2 are activated The displayed variable is represented by an activated single bar which moves between the low ...

Страница 27: ...xceed those specified by clause 5 7 of EN 60079 11 for simple apparatus which simplifies system design and documentation Providing the increased voltage drop can be tolerated the intrinsic safety and system design described in sections 3 and 4 of this manual remain valid with the loop powered backlight 9 4 2 Separately powering the backlight The optional backlight may also be powered from a separa...

Страница 28: ... is likely to occur occasionally in normal operation Zone 22 explosive atmosphere in the form of a cloud of combustible dust in air is not likely to occur in normal operation but if it does occur will only persist for a short period Be used with dust in subdivisions IIIA combustible flyings IIIB non conductive dust IIIC conductive dust For use with IIIC conductive dusts special conditions for safe...

Страница 29: ...he inside of an Ex t enclosure is effectively a safe area and the certified intrinsically safe interface may therefore be mounted in the same Ex t panel enclosure as the indicator The indicator terminals the wiring to the indicator and the intrinsically safe interface if mounted within the enclosure should be segregated from all other non intrinsically safe wiring and equipment within the panel en...

Страница 30: ...of this manual may therefore be used for IECEx installations but the local code of practice should also be consulted A2 3 Special conditions for use in IIIC dusts The IECEx intrinsic safety certificate has an X suffix indicating that for some applications special conditions apply for safe use which are identical to ATEX special conditions for safe use a When installed in an Ex px Ex py or Ex pz pa...

Страница 31: ... parameters are similar to the ATEX and IECEx parameters and the systems shown in section 4 of this manual may therefore also be used for FM and cFM installations The FM and cFM Control Drawing CI300 72 for intrinsically safe applications which is contained in this appendix specifies that when correctly installed a BA307E SS or a BA327E SS will not invalidate the certification of an AEx e AEx p or...

Страница 32: ...32 ...

Страница 33: ...33 ...

Страница 34: ...34 ...

Страница 35: ...35 ...

Страница 36: ...36 ...

Страница 37: ...37 ...

Страница 38: ...38 ...

Страница 39: ...39 ...

Страница 40: ...40 ...

Страница 41: ...41 ...

Страница 42: ...42 ...

Отзывы: