Cleaning / Disinfection 

User Manual AVERO Premium Plus V2.4 E 

 

Seite 

37

 von 

58

 

11.  Cleaning / Disinfection 

Cleaning the AVERO Premium Plus Lifting Tub 

  Clean the AVERO Premium Plus lifting tub with a soft, lint-free cloth. 

  Use a cloth dampened with soapy water for thorough cleaning.  

  For disinfection, a cloth moistened with an isopropyl alcohol solution can be used. 

  To avoid damage, 

do not use

 aerosol cleaners, sprays, abrasives or solvents to clean 

the control panel. 

 

Attention:

 

Make sure that the system is not in operation during 

cleaning!

 

 
 

Warning:

 

After each use, complete disinfection of the AVERO 
Premium Plus with a disinfectant must be performed in 
order to avoid cross-contamination. 

 
The AVERO Premium Plus lifting tub can be cleaned while damp with a standard plastic 
cleaner.  

To disinfect the surface, use an isopropyl alcohol solution or an industry-standard disinfectant 
aerosol (spray).  

The tub body of the lifting tub AVERO Premium Plus consists of fiber glass, reinforced plastic 
and is to be disinfected with surface disinfectant (not aggressively). 

If aggressive cleaning agents are still used, all warranty claims regarding the surface damage 
are lost. 

 

Disinfecting the AVERO Premium Plus Lifting Tub 

With regard to the transfer risks, we recommend that carefully you disinfect and rinse your bath 
between each use. 

The disinfection of the tub is carried out manually with the disinfection handheld showerhead 
(yellow). 

Please note that you can only disinfect with a disinfectant system, not clean. 

To remove grease stains, skin debris and hair, use a sponge and soap. 

Never use abrasives to clean the tub! 

 

 

Содержание AVERO Premium Plus

Страница 1: ...BEKA Hospitec Ltd Edition V2 4 E Subject to technical modifications 2018 02 14 User Manual for AVERO Premium Plus Lifting Tub...

Страница 2: ...BEKA Hospitec Ltd AVERO Premium Plus User Manual V2 4 E Page Page 2 of 58...

Страница 3: ...ents of this user manual before operating the unit Please note the safety instructions in particular Slight deviations of the actual delivered units from the pictures shown here are possible If you ha...

Страница 4: ...Premium Plus Lifting Tub 20 Application areas 20 Duration of use 20 Typical indications 20 Contraindications 20 6 Safety Information 21 7 AVERO Premium Plus Lifting Tub Operating Elements 24 7 1 Oper...

Страница 5: ...ackaging waste 39 Disposal of product 39 13 Applicable test procedures controls 40 14 Troubleshooting and customer service 41 Help Troubleshooting 41 Customer service 42 15 Appendix 43 15 1 Technical...

Страница 6: ...by BEKA Hospitec In case of technical interventions such as additions or modifications to our equipment which are not carried out by the BEKA Hospitec or by BEKA Hospitec authorized specialist compan...

Страница 7: ...g to the EC Directive concerning medical devices can only be used indoors Protection class I IPX4 Protection against penetration of liquids Protection against splashing water on all sides Type label P...

Страница 8: ...installation site Use a pallet or similar for transporting Arrange the tub so that it is secure and cannot tip Tools Required To remove the packaging you need Sharp bladed cutter Remove the cardboard...

Страница 9: ...asteners are released you can remove the lifting tub from the pallet and position it in the intended place of use Remove the bubble wrap and stretch film Make sure that you do not damage the surface o...

Страница 10: ...is device is not approved for operation in potentially explosive atmospheres Maintenance of the unit must be carried out only by us or by a party expressly authorized by us otherwise the warranty shal...

Страница 11: ...local water supply may require readjustment of the thermostatic mixer 4 2 Electrical system Connection to the power grid When connecting the device to an electric power grid AC Power supply AC outlet...

Страница 12: ...d tilting of the tub the tub has to be additionally secured Note they refer to the following pictures Illustration 1 Illustration 2 Illustration 3 Illustration 4 Illustration 5 Illustration 6 Illustra...

Страница 13: ...RO Premium Plus 175 Wall installation Illustration 1 Illustration 2 Illustration 3 Illustration 4 Illustration 5 Illustration 6 Illustration 7 Illustration 8 Illustration 9 Illustration 10 Illustratio...

Страница 14: ...Unpacking and assembly User Manual AVERO Premium Plus V2 4 E Seite 14 von 58 Illustration 17 Components 18 Tools18...

Страница 15: ...ss screw from the rotary actuator and pull it forwards Figure 5 Open a solenoid valve to check the water temperature In the pictures shown handbrake Figure 6 Check the outlet temperature with the exte...

Страница 16: ...te If the temperature is set at 42 C anti scalding the corresponding display HOT alternates with the temperature and the water flow is switched off The handshower can then be switched on again with th...

Страница 17: ...Turn the knob to the left until it stops Snap button not pressed Figure 3 Use the screwdriver to remove the cap Figure 4 Unscrew the cross screw from the rotary actuator and pull it forwards Figure 5...

Страница 18: ...gure 11 Activate the thermal flushing for the water inlet Press and hold button 6 than press the buttons 8 and 10 together Perform the thermal rinse of the water spout according to the valid laws of y...

Страница 19: ...The picture shows the desired maximum temperature which must agree with the manually measured temperature Figure 17 Slide the knob to the maximum position Figure 18 Switch off the solenoid valve Figur...

Страница 20: ...itals nursing homes facilities for the disabled and private households Contraindications WARNING Patients with Heart disease arrhythmias Tumors in the application area Leukemia Bacterial infection Rhe...

Страница 21: ...t to the tub before and during the height adjustment The operator must ensure that there is nothing in the area above and below the tub which could impede the free upward and downward movement of the...

Страница 22: ...be damaged A damaged power cord may cause fire or a life threatening electric shock and may not be used Customer service If the device is defective and none of the measures described in section 13 he...

Страница 23: ...ut in conjunction with maintenance The intervals are defined in Section 13 Accessories Operate the AVERO Premium Plus lifting tub only with original accessories WARNING Any unauthorized repairs modifi...

Страница 24: ...2 Bath shortener 7 USB port for music sound system MP3 3 Eccentric rotary operation 8 Handheld showerhead 4 Bath spout 9 Thermostatic mixer with scald protection 5 Control panel with tub functions 10...

Страница 25: ...g User Manual AVERO Premium Plus V2 4 E Seite 25 von 58 Picture 4 Picture 5 Temperature sensor for the bath water temperature Eccentric stopper with tub drain Eccentric rotary operation Thermostat Air...

Страница 26: ...rature display 10 Color light therapy system button 3 Filling temperature display 11 Jet button 4 Bathwater temperature display 12 Height adjustment button UP 5 Handheld showerhead button 13 Height ad...

Страница 27: ...y pressing the Button again Note Check the water temperature by hand without gloves or with a thermometer before the patient steps in the tub Wheelchair patients can easily be lifted with mobile lift...

Страница 28: ...s A disinfection procedure must last for at least 15 seconds otherwise the system starts from a faulty operation The disinfection can be activated manually by pressing the DIS key 1 or end automatical...

Страница 29: ...the Handheld showerhead 42 C is greater scald protection display flashes HOT If the scald protection activates temperature display flashes HOT the handheld showerhead cannot be turned on using the bu...

Страница 30: ...ium intensity With button the power of the fan can be adjusted accordingly The controller also has an interval function for the fan The interval function is activated by simultaneously pressing the in...

Страница 31: ...tart the color change again To turn off the color light therapy press the button 10 for min 3 seconds 10 7 Power Function LED Display 14 The LED display 11 shows you the status of the bathtub The foll...

Страница 32: ...Operating User Manual AVERO Premium Plus V2 4 E Seite 32 von 58 10 8 Music Sound system...

Страница 33: ...turned to ON The switch for activating the emergency drain function is located under the removable inspection cover for refilling the disinfectant Note The battery capacity is designed so that you can...

Страница 34: ...en the tub has regained sufficient voltage the controller will revert to a normal state All displays will be switched to their normal state and the internal batteries will be reloaded to their intelli...

Страница 35: ...nderful experience for the senses of your resident An easy operation and a maintenance free system are self evident This option is only possible in combination with the air jet whirl system Filling th...

Страница 36: ...life is dependent on how often and how long the application lasts Usually this is 3 months Nevertheless one should regularly check whether the indicated flavors are still sufficient CAUTION Do not pou...

Страница 37: ...Plus lifting tub can be cleaned while damp with a standard plastic cleaner To disinfect the surface use an isopropyl alcohol solution or an industry standard disinfectant aerosol spray The tub body of...

Страница 38: ...ection cover on the back side of the AVERO Premium Plus lifting tub When refilling and replacing the container make sure that the hose reaches the bottom of the container 2 Take the disinfection handh...

Страница 39: ...with the AVERO Premium Plus lifting tub according to local regulations and laws for recycling The metal parts and plastic and electronic components should be recycled or disposed of in accordance wit...

Страница 40: ...Rule 3 BGV A3 test on the AVERO Premium Plus lifting tub is performed annually This DGUV test only is required for Germany In other countries other rules may apply Daily before each use Clean disinfe...

Страница 41: ...k filters in the water inlet Autofill not working Tub power switched off Performing a restart If this does not achieve the desired result inform customer service Only hot or cold water is running from...

Страница 42: ...k c The remote control batteries need to be replaced The color light therapy does not work a Mainboard defective Inform customer service b Headlamp defective Inform customer service Unusual noises whe...

Страница 43: ...90 approx 250 kg Weight without water 170 approx 230 kg Capacity 190 without patient approx 220 l Capacity 170 without patient approx 190 l Patient weight max 220 kg 485 lbs SWL Safe load 300 kg 661 l...

Страница 44: ...t Accuracy of the display 1 C Environmental conditions Operation Temperature range 10 C to 40 C Relative humidity 30 to 95 non condensing Atmospheric Pressure 800 1060 hPA Storage and transport Temper...

Страница 45: ...Appendix AVERO Premium Plus User Manual V2 4 E Page Page 45 of 58 15 2 Dimensions AVERO Premium Plus 190...

Страница 46: ...Appendix AVERO Premium Plus User Manual V2 4 E Page Page 46 of 58 15 3 Dimensions AVERO Premium Plus 170...

Страница 47: ...Appendix AVERO Premium Plus User Manual V2 4 E Page Page 47 of 58 15 4 Installation plan AVERO Premium or 110V 60Hz...

Страница 48: ...Appendix AVERO Premium Plus User Manual V2 4 E Page Page 48 of 58 15 5 Declaration of Conformity...

Страница 49: ...allation may require the expertise of a trained specialist in electrical plumbing Article No Description 910131100 Bath shortener 910131203 Cleaning system 910131300 Automatic water stop 910131403 Mus...

Страница 50: ...NT specified below The customer or user of the AVERO Premium Plus should ensure that it is used in such an environment Release measurements Agreement Electromagnetic environment guidelines High freque...

Страница 51: ...e for input output lines The quality of the power supply should be that of a typical commercial or hospital environment Overload IEC 61000 4 5 to 1 kV line opposite line to 2 kV line opposite ground t...

Страница 52: ...ters by electromagnetic location measurement a should be less than the CONCORD level in all frequency ranges b In the vicinity of equipment marked with the following symbol interferences are possible...

Страница 53: ...f the device W Safe distance depending on transmission frequency in m 150 kHz to 800 MHz d 0 35 P 80 MHz to 800 MHz d 1 2 P 800 MHz to 2 5 GHz d 2 3 P 0 01 0 1 1 10 100 0 04 0 11 0 35 1 1 3 5 0 12 0 3...

Страница 54: ...82 Wetzlar Serial Number _____________________________________________ Purchase date _____________________________________________ Location _____________________________________________ Tests carried...

Страница 55: ...Appendix AVERO Premium Plus User Manual V2 4 E Page Page 55 of 58 Repeat examination repair DGUV 3 Technical Safety Check etc Type of examination Date Result Action Signature...

Страница 56: ...__________ ____________________________________________________________________________________________________ ________________________________________________________________________________________...

Страница 57: ...__________ ____________________________________________________________________________________________________ ________________________________________________________________________________________...

Страница 58: ...BEKA Hospitec Ltd Am R benmorgen 3 D 35582 Wetzlar Telephone 49 0 6 41 9 22 20 0 Fax 49 0 6 41 9 22 20 20...

Отзывы: