Operation
Operating manual CARLO EP & Classic
V2.10 E
Page 21 of 56
7.10 Motor Safety Measures
The control of the electrical motor is equipped with an overload protection, which is
autonomously switched off in case of overload. The motor will only be operational
again after a short waiting time. The cooling-off time of the motor can be up to 18
minutes depending on the ambient temperature.
NOTE:
Opening the motor will result in expiration of the
warranty.
The intermittent operation of the motor is not a defect,
but is only for your own safety.
7.11 Impact and Jamming Protection (hoist motor)
The electrical motor (hoist motor) features an integrated impact and jamming
protection (freewheeling). This feature avoids jamming, pinching and/or crushing
dangers when the spreader bar of the CARLO ALU strikes or encounters an obstacle.
The motor runs free until the obstacle is removed or the CARLO ALU is moved away
from the obstacle.
After the obstacle has been removed, the spreader bar could start lowering
autonomously. Therefore, you must immediately release the button of the handset as
soon as you have noticed that an obstacle has been encountered.
8 Operation
8.1 Sling Operating Manual
Important note:
The normal service life of BEKA-slings is approx. 36 months from the date of
production (marked on the sling). The specified service life is only applicable, when the
BEKA-slings are cleaned, maintained and inspected in accordance with the
instructions contained in the following documentation.
8.1.1 Prior to Use
The slings must be checked before and after each use and, if necessary, be washed in
accordance with the manual. This is prescribed in particular to reduce the risk of
infections to an absolute minimum, in the event that the same equipment is used for
other residents or patients.
Prior to each use, a thorough check of the sling including the loops and the fastening
clips is imperatively required. If the sling or the loops would be frayed, cut-in or
damaged, or the clips would be damaged, the sling no longer must be used.