BEING Scientifi c Inc.
User
Edition
15
Changing Power-up Mode
Units are shipped with power-up mode set to 2. The
controller starts running from the last power-off.
Parameter
Po=0002
.
STEP 1:
Enter Level 2 code. LK=0003
Press the INCREASE key to enter the code digit.
Press the SETTING key to enter the Level 2 controller
parameters.
SET
C
C
TEMP
2 0
22
63
4
0
50
Blink
1
time
STEP 2:
Scroll to parameter Po
The tM parameter will appear. Press the SETTING
key 1 time to scroll through parameters until the
parameter Po appears.
STEP 3:
Change parameter Po
Press the DECREASE key to change the fi rst
digit.
Po=0000 The controller is in a stopped
state after power-on, and user must start the
operation pressing START/STOP button for 2
seconds.
Po=0001 The controller automatically the step
operation after power-on.
Po=0002 The controller starts running from
the last power-off.
Press the SETTING key 4 times to set parameter Po
and return to standard state.
STEP 4:
Set parameter and return to standard state.