Controller used on the following product

Ÿ

BOF/BON Drying Ovens

Ÿ

BIT/BIF Heating Incubators

Ÿ

BIC Cooling Incubators

Ÿ

BOV Vacuum

vens

O

Ÿ

BWB Water Bath

User Edition

Содержание L Series

Страница 1: ...R L Series PROG STEP TIME SET RUN C C TEMP 88 88 88 888 888 8 8 88 UV H L F F Controller used on the following product BOF BON Drying Ovens BIT BIF Heating Incubators BIC Cooling Incubators BOV Vacuum vens O BWB Water Bath User Edition ...

Страница 2: ... 120 200 400 Mechanical Convention Drying Ovens BON 30 50 115 200 Natural Convention Drying Ovens BOV 20 50 90 Vacuum Ovens BWB 05 12 22 General Purpose Water Bath 2 0 TABLE OF CONTENTS Safety Messages 03 Controller elements overview 04 Changing temperature unit of measurement 05 Calibration after unit of measurement change 06 07 What is a multi step program 08 Turn on off multi step programs 09 P...

Страница 3: ... operates this equipment at all times Keep non qualified personnel at a safe distance from this unit Only qualified personnel familiar with the safe installation operation and maintenance of this unit should attempt start up or operating procedures Always stop the instrument before making or removing any connections Symbols used in this Programming Guide The following signal word panels safety symbo...

Страница 4: ... when heating elements are off COOLING Snowflake illuminates and blinks when the unit s cooling is on Turns off when cooling system is off Series BIC only FAN Fan blades illuminate and blinks when the unit s circulation fan is on H high M medium and L low indicate the fan speed Turns off when fan is off Series BIF BIC and BOF only Water Level Indicates high and low water bath levels Visible Alarm I...

Страница 5: ...e Level 3 controller parameters First available parameter will appear Press the SETTING key 15 times to scroll through parameters until the parameter CF appears Press the INCREASE or DECREASE key to change the first digit CF 0000 Celsius O C CF 0001 Fahrenheit O F Press the SETTING key 2 times to set parameter CF and return to standard state Upon changing the temperature unit of measure ment the pr...

Страница 6: ...t 10O C 18O F greater than the first STEP 4 Compare thermometer readings to set and present temperatures If SV1 PV1 TV1 or SV2 PV2 TV2 move to STEP 5 STEP 1 Place test materials in chamber set program and press run for 3 seconds The test hole accessory port is Ø25mm and located on the right side of the incubator Upon reaching the set temperature and opening the test hole insert thermometer probe in...

Страница 7: ...ming Correction Press the SETTING and SHIFT keys for 3 seconds to enter the controller s internal parameter settings Time area will change and first digit will blink SET C C TEMP 00 22 63 4 0 00 Blink STEP 6 7 Enter Level 2 code LK 0003 STEP 7 8 Change parameter Pb Press the INCREASE key to enter the code digit Press the SETTING key to enter the Level 2 controller parameters The tM parameter will a...

Страница 8: ... time runs out unless it is the last step in the program then the unit shuts down and the audible and visual alarm goes off Numerous applications like materials testing quality assurance environmental simulation and many more in many industries require precise multi temperature heating or cooling testing BEING s L Series controller allows the user to easily program up to 8 multi step multi tempera...

Страница 9: ...code LK 8286 STEP 3 Change parameter Mo STEP 4 Set parameter and return to standard state Press the INCREASE key to enter the first code digit Press the SHIFT key to move to the second code digit Digit will blink Press the INCREASE key to enter the second code digit Follow the same procedure for the third and fourth code digits Press the SETTING key to enter the Level 4 controller parameters The Mo...

Страница 10: ...are turned on If multi step programs are turned off and later turned on the last program run or selected will be displayed Press the SETTING key for 3 seconds The PROG 1 will initially blink Then the number will blink Units are shipped with all steps in each program set to zero temperature and zero run time Pressing the SETTING key starts the programming process STEP 1 will appear End in TIME area...

Страница 11: ...p programs are turned off and later turned on the last program run or selected will be displayed Press the SETTING key one 1 time The PROG area will blink STEP 2 Select program number Press the INCREASE or DECREASE to select program number Press the SETTING key to confirm the pro gram to be run Pressing the RUN key for 2 seconds starts the program STEP area RUN HEAT and run time illuminate STEP 3 P...

Страница 12: ...el 1 parameters Press the INCREASE key to enter the first cycle value Press the SHIFT key to move to the second cycle digit Digit will blink Press the INCREASE key to enter the second cycle value Press the SETTING key 1 time to scroll through parameters until the parameter Cy appears STEP 2 Select program 1 8 to run multiple cycles See Selecting Multi step Program to Run instructions on page 11 STE...

Страница 13: ...blink STEP 2 Press SETTING key Press the SETTING key to enter Level 1 parameters STEP 1 Enter Level 1 of controller internal parameters When entering the controller s internal parameters Level 1 LK 0000 will always come up first Press the SETTING key 1 time to set parameter dy and return to fixed value program standard state Press the INCREASE key to enter the first time value Press the SHIFT key to ...

Страница 14: ...git will blink STEP 2 Press SETTING key Press the SETTING key to enter Level 1 parameters STEP 2 Select program 1 8 to run multiple cycles See Selecting Multi step Program to Run instructions on page 11 STEP 3 Enter Level 1 of controller internal parameters When entering the controller s internal parameters Level 1 LK 0000 will always come up first SET C C TEMP 222 0 4 0 PROG 000 1 Blink 2 times SE...

Страница 15: ... Po The tM parameter will appear Press the SETTING key 1 time to scroll through parameters until the parameter Po appears STEP 3 Change parameter Po Press the DECREASE key to change the first digit Po 0000 The controller is in a stopped state after power on and user must start the operation pressing START STOP button for 2 seconds Po 0001 The controller automatically the step operation after power ...

Страница 16: ...function when the temperature exceeds the Setpoint AL value Pb 000 0 Zero Adjustment Intercept 100 0 100 0 When the zero error of the meter is large and the full scale error is small adjust this value Generally PT100 rarely adjusts this value PK 0000 Full Scale Adjustment Slope 1000 1000 When the zero error of the meter is small and the full scale error is large adjust the value PK 4000 x mercury ...

Страница 17: ...cientific Inc User Edition PROMPT INITIAL VALUE FUNCTION NAME FUNCTION RANGE EXPLANATION Mo 0000 Program Type Selection 0 1 Mo 0000 Fixed Value Single step Program Mo 0001 Multi step Program Level 4 LK 8286 17 ...

Страница 18: ...BEING Scientific Inc User Edition 18 INTENTIONALLY BLANK ...

Страница 19: ...BEING Scientific Inc User Edition 19 INTENTIONALLY BLANK ...