52
5.2 Capacity Measurement
Prior to any capacitor test, it
must be ensured, that the
capacitor to be tested is com-
pletely discharged. Failure to
comply with this prescription
can lead to dangerous corpo-
ral user injuries or cause
instrument damage. Additio-
nally, charged capacitors fal-
sify the measurement result.
Charged capacitors have to be
discharged using a high-ohm
resistance (not via short-
circuit !).
When dealing with polarised
capacitors, connect ”-” pole
to the multimeter COM
socket.
Z
Resistors and semiconductor
paths in parallel to the capacitor
cause falsified measurement
results.
UNITEST Hexagon 330 and
UNITEST Hexagon 340:
❿
Select measurement range
via measurement function selec-
tion switch.
❿
Connect the black test lead to
the COM socket and the red
test lead to the Temp/
/
/
socket.
❿
The capacity test is pre-set. If
the diode symbol appears on
the display the key
is
used to switch to capacity test.
UNITEST Hexagon 310 and
UNITEST Hexagon 320:
❿
Select measurement range
via measurement function selec-
tion switch.
❿
Connect the black test lead to
the COM socket and the red
test lead to the
/
/
socket.
❿
Connect test leads to UUT.
❿
Read the measurement result
displayed on the screen.
Capacity Measurement