Carrying out Measurements
6.4 AC Voltage Measurement
Never apply voltages exceeding 300V to input
sockets.
Connect mains lead (13) to the mains
connector
Select function UAC via the function switch
(17)
Select the contact voltage by means of key
(15) in display
Connect the the test leads /mains plug to
UUT
Read the measurement result on the display
If the voltage exceeds 300V, the message "At-
tention (1) + Limit (2) " is displayed. In this in-
stance, immediately disconnect the measure-
ment instrument from UUT.
Also refer to chapter 7.0.
No possibility to store the measurement re-
sults.
6.5 Frequency Measurement
Never apply tvoltages exceeding 300V to input
sockets.
Connect the mains lead to the mains connec-
tor (13)
Select the function f via the function switch
(17)
Connect the the test leads /mains plug to UUT
Read the measurement result on the display
Also refer to chapter 7.0.
No possibility to store the measurement
results.
6.6 Contact Electrode
Measurement of the present contact voltage is car-
ried out after setting the contact voltage limit. By
touching the contact electrode, exceeding the limit
is automatically indicated (>25V or 50V).