background image

  

  

  

3  

  

9.

 

On  the  piCO

  and  piCO

baby

,  when  the  test  is  finished  

  

  will  appear  at  the  bottom  

of  the  screen.  

  

10.

 

On   the   Micro

+

™,   when   the   test   is   finished     

  

  

   will   appear   at   the   bottom   of   the  

screen.  

  

11.

 

If  a  high  reading  has  been  recorded,  you  can  mute  the  sound  by  pressing  

.  

  

12.

 

To  repeat  breath  test,  press  

  once  to  return  to  the  home  screen  and  repeat  steps  3-­‐8.  

  

13.

 

To  save  the  reading  (Micro

+

™  only)  press  

  and  select  the  relevant  patient  profile.  

  

14.

 

Remove  the  D-­‐piece  between  tests  to  purge  sensor  with  fresh  air.  

  

15.

 

To  switch  off,  press  and  hold  the  power  button  for  3  seconds,  unit  will  also  power  off  after  2  
minutes  of  inactivity  to  save  power.    

  

  
Conducting  an  Ambient  Test  –  Micro

+

™  only    

  
The   environmental   air   can   be   tested   to   check   for   CO   in   the   environment   using   the   Micro

+

™  

Smokerlyzer.  This  is  particularly  useful  in  the  event  of  a  patient  reporting  a  non-­‐smoker  status  but  
providing  a  breath  reading  of  that  of  a  smoker  suggesting  they  have  inhaled  environmental  CO.    
  

This   is   carried   out   by   pressing  

   on   the   screen;   the   Micro

+

™   will   then   carry   out   real-­‐time  

monitoring  of  the  environmental  air.  To  stop  the  test,  press  

.  

  
Bedfont   do   not   recommend   that   the   environmental   test   alone   is   used   as   a   measure   of   passive  
smoking  as  the  very  high  dispersiveness  of  CO  at  the  levels  involved  with  smoking  can  lead  to  false  
negative  readings.  
  

Reviewing  history  –  Micro

+

™  only    

  

The  Micro

+

™  will  record  every  reading  taken  in  its  history  up  to  150  readings,  to  access  this  press  

  

and  then  

  the  history  will  then  be  shown.  

  

Changing  date  and  time  –  Micro

+

™  only    

  

Press  

  and  then  

  you  can  then  select  either  d-­‐m-­‐y  or  m-­‐d-­‐y  for  the  date  format  and  12h  or  

24h  for  the  time  format.  Dates  and  times  are  then  adjusted  by  selecting  the  number  you  wish  to  

change  and  pressing  

or  

  to  increase  and  decrease.  Press  

  to  save  your  settings.  

  

  

Содержание Smokerlyzer Range

Страница 1: ...Smokerlyzer Range For use with piCO piCObaby and Micro Scientific contributions to health User manual ...

Страница 2: ... and time Micro only 3 Traffic Light Pre set Cut Off Levels 4 Technical Specification 4 Safety Information and Device Symbols 4 Environment 5 Instrument Layout 5 Display and Device Symbols 6 Warnings and Maintenance 8 Cleaning 8 Routine Maintenance 9 Troubleshooting 9 COdata Software connecting to the PC 10 Returns Procedure 11 Spares 11 Warranty 11 ...

Страница 3: ...has been inhaled and the COHb and fCOHb readings relate to the percentage of vital oxygen that has been replaced in the bloodstream The piCO Smokerlyzer displays both CO ppm and COHb the piCObaby Smokerlyzer displays both CO ppm FCOHb and COHb and the Micro Smokerlyzer displays CO ppm COHb and FCOHb However all devices only directly measure CO ppm the COHb and fCOHb is a calculation based on clini...

Страница 4: ...ly useful in the event of a patient reporting a non smoker status but providing a breath reading of that of a smoker suggesting they have inhaled environmental CO This is carried out by pressing on the screen the Micro will then carry out real time monitoring of the environmental air To stop the test press Bedfont do not recommend that the environmental test alone is used as a measure of passive s...

Страница 5: ...es on new alkaline batteries approx 3 min per use Approx 1 x CR2032 Battery Life 175 000 hours T90 Response Time 30 seconds Operating Temperature Range 0 40 C storage 0 50 C Operating Transport Storage Pressure 800 1200mbar Operating Humidity 10 90 non condensing storage 0 95 Sensor Operating Life 5 years 5 year warranty Sensor Sensitivity 1ppm Sensor Drift 5 per annum Dimensions Approx 34 x 75 x ...

Страница 6: ... by cellular phones and by electromagnetic interference exceeding the levels specified in EN50082 1 This equipment should be moved if necessary to avoid interference Instrument Layout 1 Display 2 Power button 3 Breath sampling D Piece 4 D Piece aperture 5 Exhaust port for breath sample 6 USB connector for use with COdata Software 7 Single use SteriBreath mouthpiece 8 Battery compartment 9 Battery ...

Страница 7: ...Micro Battery condition Full Battery condition Low Battery condition Empty Breath Test Environmental Test N A N A Settings Patient Profiles N A N A View patient results N A N A Inhale Hold Breath Countdown Timer Exhale ppm Reading COHb Reading Show fCOHb N A ...

Страница 8: ... fCOHb Reading N A Ambient CO reading N A N A Home Change D Piece Change Traffic Light Cut off Levels Save N A N A Change Breath Hold Time N A N A History N A N A Mute Next Step Countdown to Sensor Change Change Sensor Calibrate Device ...

Страница 9: ...ning to appropriately qualified persons 5 Holding the reset button down for 30 seconds will perform complete device reset this will clear any saved data and revert all settings to the factory defaults After performing a reset the device will need to have the date time set and be calibrated before it can be used Cleaning 1 The Smokerlyzer products are moulded with SteriTouch technology for optimum ...

Страница 10: ...mbols moulded into the plastic High readings recorded for reported non smoker If a breath test on a non smoker gives an indication of CO being present it may indicate the following 1 A high ambient level of CO possibly due to a faulty combustion appliance or exhaust fumes 2 The effects of passive smoking 3 The device is contaminated by alcohol or another organic substance 4 The patient may have a ...

Страница 11: ... be shown onscreen Follow the onscreen instructions to finish the calibration process 6 Gas flow should be set to 1 0 litres per minute 7 A successful calibration will be indicated by the press to return to the home screen 8 A failed calibration will be indicated by press to attempt calibration again If the problem persists see Troubleshooting or call your local supplier of Bedfont products 9 Retu...

Страница 12: ...Department if you are unable to supply an Official Purchase Order Number 5 If your monitor is still in warranty and the fault is covered by warranty see Warranty section of this manual Bedfont will repair it and return it to you with an Engineer s Report free of charge 6 If you choose not to proceed with the repair a handling fee will be charged Ensure that you return the completed authorisation f...

Страница 13: ... heure Seulement Micro 14 Classification du fumeur et code couleur prédéfini 15 Spécifications techniques 15 Informations de sécurité des icônes 15 Environnement 16 Schéma d appareil 17 Affichage et icônes d appareil 17 Avertissements et maintenance 20 Nettoyage 20 Entretien courant 21 Dépannage 21 Logiciel COdata branchement au PC 22 Procédure de retour 22 Pièces consommables 23 Garantie 23 ...

Страница 14: ...té de gaz dans les poumons les voies respiratoires correspondant à la quantité de CO nocif qui a été inhalée et la COHb et fCOHb à la quantité de gaz présente dans le sang L analyseur piCO Smokerlyzer affiche la concentration en CO ainsi que COHb Le piCObaby Smokerlyzer et le Micro Smokerlyzer affichent en plus fCOHb Cependant tous ces appareils ne mesurent que directement ppm CO la COHb et fCOHb ...

Страница 15: ...on enfoncé pendant 3 secondes L analyseur s éteindra automatiquement après 2 minutes d inactivité Effectuer un test ambiant Seulement Micro Avec le Micro Smokerlyzer l air ambiant peut être analysé pour vérifier la concentration de CO dans l environnement Ceci est particulièrement utile dans le cas où un patient qui est signalé comme étant non fumeur et obtient un résultat correspondant à celui d ...

Страница 16: ...modifiées Spécifications techniques piCO piCObaby Micro Plage de concentration CO 0 150ppm 0 150ppm 0 500ppm Affichage Ecran tactile couleur Principe de détection Capteur électrochimique Répétabilité 3 sur lectures consécutives Interférence avec H2 6 Alimentation électrique 3 x AA LR6 ou équivalent 1 x CR2032 pile au lithium Autonomie avec 3 piles AA 1 000 utilisations avec des piles alcalines neu...

Страница 17: ...ce des mélanges inflammables Référer aux instructions pour l utilisation Courant continu Marquage CE Pièce appliquée de type BF Eliminer selon DEEE Numéro de série Fabricant et date de fabrication Logo de Bedfont Environnement Le piCO piCObaby et le Micro Smokerlyzer sont conformes à la norme EN60601 1 2 sur la compatibilité électromagnétique mais peuvent être affectés par les téléphones portables...

Страница 18: ... pour utilisation avec le logiciel COdata 7 Embout buccal SteriBreath à usage unique 8 Compartiment des piles 9 Clip de compartiment des piles 10 Bouton de réinitialisation 11 Interrupteur de programmation Affichage et icônes d appareil Description piCO piCObaby Micro Etat des pile Pleine Etat des piles Faible Etat des piles Vide Test respiratoire Test ambiant N A N A Paramètres ...

Страница 19: ...Profils des patients N A N A Afficher les résultats mémorisés N A N A Inspirer Retenir la respiration Compte à rebours Expirer doucement et complètement Lecture en ppm Lecture en COHb Afficher la fCOHb N A ...

Страница 20: ... N A Accueil Remplacer le piégeur Dpiece Modifier les seuils de classification du fumeur Enregistrer N A N A Modifier le compte à rebours pour le temps d apnée N A N A Historique N A N A Volume Prochaine étape Date de remplacement du capteur Remplacer le capteur Etalonner l appareil ...

Страница 21: ...u dispositif 10 En maintenant le bouton de réinitialisation pendant 30 secondes le dispositif se réinitialisera complétement ceci effacera toutes les données enregistrées et rétablira tous les paramètres aux valeurs par défaut Après une réinitialisation il faut vérifier l heure et la date et étalonner le dispositif avant utilisation Nettoyage 4 Les produits de la gamme Smokerlyzer sont moulés avec...

Страница 22: ...es Assurez vous que les piles sont bien placées dans le bon sens dans le compartiment Résultats élevés pour un non fumeur Si un non fumeur obtient des résultats élevés lors du test respiratoire les explications sont les suivants 6 Un niveau élevé de CO ambiant peut être dû à un appareil à combustion défectueux ou à des gaz d échappement 7 Les effets du tabagisme passif 8 L appareil est contaminé p...

Страница 23: ...r revenir à l écran d accueil 17 Si l étalonnage a échoué l écran affichera cette icône appuyez sur l icône pour réessayer l étalonnage Si le problème persiste voir Dépannage ou appelez votre revendeur local 18 Appuyez sur l icône pour revenir à l écran d accueil Logiciel COdata branchement au PC Branchez le câble dans le port USB de l ordinateur et le port de connexion USB en haut de l appareil A...

Страница 24: ...iel une fois que Bedfont a reçu les documents demandés Si l opération est hors garantie des frais de livraison seront facturés Pièces consommables D PIECE 2 Boîte de 12 piégeurs Dpiece non stérilisable STERIBREATH Boîte de 250 embouts buccaux SteriBreath non stérilisable WIPE V Paquet de 50 lingettes de nettoyage FXS397 Câble USB Garantie Bedfont Scientific Limited garantit les piCO piCObaby Micro...

Страница 25: ...e without prior notice Registered in England and Wales Registered No 1289798 Bedfont Scientific Limited 2012 ISO 9001 2008 Cert No FM 31664 ISO 13485 2003 Cert No MD 502905 Contact Bedfont or one of our worldwide Smokerlyzer distributors for a free demonstration www bedfont com Tel 44 0 1622 851122 Email ask bedfont com A full list of our worldwide distributors can be found at http www bedfont com...

Отзывы: