23
Guidance and manufacturer’s declaration: Electromagnetic Immunity (IEC 60601
-1-2)
The Smokerlyzer® monitors are intended for the use in the electromagnetic environment specified below. The
customer or the user of the Smokerlyzer® should assure that it is used in such an environment.
Emissions test
IEC 60601 Test level
Compliance level
Electromagnetic
environment
–
guidance
Radiated Emissions
EN55011:2007
30MHz to 1GHz
Met or exceeded
None
Immunity test
IEC 60601 Test level
Compliance level
Electromagnetic
environment
–
guidance
Radiated Immunity
EN61000-4-3:2006 +
A1:2008 + IS1 2009 +
A2:2010
10V/m (1kHz 80%) 80MHz
–
2.7GHz
385 MHz 27 V/m PM 18 Hz
450 MHz 28 V/m FM 1 kHz sine
710 MHz 9 V/m PM 217 Hz
745 MHz 9 V/m PM 217 Hz
780 MHz 9 V/m PM 217 Hz
810 MHz 28 V/m PM 18 Hz
870 MHz 28 V/m PM 18 Hz
930 MHz 28 V/m PM 18 Hz
1720MHz 28 V/m PM 217 Hz
1845 MHz 28 V/m PM 217 Hz
1970 MHz 28 V/m PM 217 Hz
2450 MHz 28 V/m PM 217 Hz
5240 MHz 9 V/m PM 217 Hz
5500 MHz 9 V/m PM 217 Hz
5785 MHz 9 V/m PM 217 Hz
Met or exceeded
Interference may occur in
the vicinity of equipment
marked with the following
symbol:
Power Frequency Magnetic
Immunity
EN61000-4-8:2010
30 A/m magnetically sensitive
equipment
Met or exceeded
None
ESD
EN61000-4-2:1995+A1+A2
8kV contact 2, 4,8,15kV air
Met or exceeded
None