Cleaning
1.
The Smokerlyzer® products are moulded with SteriTouch® technology for optimum
infection control and bacterial efficacy. Bedfont
®
recommends wiping the instrument
and D-piece
™
external surfaces between each patient with an alcohol-free wipe
specifically designed for this purpose. A list of approved wipes can be found here:
https://www.bedfont.com/cleaning-bedfont-monitors
. The D-
piece™ cannot be
sterilised.
2.
NEVER use alcohol or cleaning agents containing alcohol or other organic solvents as
long term exposure to these vapours will damage the CO sensor inside.
3.
Under no circumstances should the instrument be immersed in liquid or splashed with
liquid
Calibration
The Smokerlyzer® monitors must be calibrated within 17-25°C. If it is too cold to calibrate (<17°C) a
blue thermometer will be shown onscreen.
Move the Smokerlyzer® to a warmer area and try again later. If it is too hot to calibrate (>25°C) a red
thermometer will be shown onscreen. Move the Smokerlyzer® to a cooler area and try again later.
Turn on the monitor by pressing
the power button once.
Press the settings icon.