Dispose of according to WEEE
Serial number
Manufacture by and date
Bedfont® logo
Environment
The Gastro
+
™ Gastrolyzer® product complies with the Medical Device Directive 93/42/EEC as
amended by 2007/47/EC and the Medical Device Electromagnetic Compatibility standard EN 60601-
1-2.
WARNING:
Portable RF communications equipment (including peripherals such as antenna cables
and external antennas) should be used no closer than 30 cm (12 inches) to any part of the Gastro
+
™
Gastrolyzer®, including cables specified by the manufacturer. Otherwise, degradation of the
performance of this equipment could result. This equipment should be moved if necessary to avoid
interference.
Guidance and manufacturer’s declaration: Electromagnetic Immunity (IEC 60601-1-2)
The Gastro
+
™
Gastrolyzer® is intended for the use in the electromagnetic environment specified below. The
customer or the user of the Gastro
+
™
Gastrolyzer® should assure that it is used in such an environment.
Test
IEC 60601 Test level
Compliance level
Electromagnetic
environment – guidance
Radiated Emissions:
FCC part 15 Subpart B
Class A & EN55011:2016 +
A1:2017 Class A
30MHz to 1GHz
Met or exceeded
Portable and mobile RF
communications
equipment should be used
no closer to any part of the
Gastro
+
™
Gastrolyzer®
including cables, than the
recommended separation
distance calculated from
the equation appropriate
to the frequency of the
transmitter.
Conducted Emissions:
FCC part 15 Subpart B
Class A & EN55011:2016 +
A1:2017 Class A
150KHz to 30 MHz
Met or exceeded
Radiated Immunity :
IEC 61000-4-3:2006 + A1
2007 + A2:2010 &
EN61000-4-3:2006 +
A1:2008 + A2:2010
3V/m (1kHz 80%) 80MHz –
2.7GHz
Spot Frequencies:
385MHz 27 V/m PM 18 Hz
450MHz 28 V/m FM 1 kHz
sine
710MHz 9 V/m PM 217 Hz
745MHz 9 V/m PM 217 Hz
780MHz 9 V/m PM 217 Hz
810MHz 28 V/m PM 18 Hz
Met or exceeded
Interference may occur in
the vicinity of equipment
marked with the following
symbol: