Medi-Gas G210
Specification
Model G210
Gas Detected
N
2
O, O
2
, CO
2
, (CO optional)
Detection Range
N
2
O 0-100%
O
2
0-100%
CO
2
0-2000 ppm
CO 0-500 ppm
Detection Resolution
Fumigation range:
1 mg/L
Sensor Type:
Internal Electrochemical Cell
Accuracy N2O Resolution: 0.1%
Accuracy: ±(1% of range + 2% of reading)
1
O2 Resolution: 0.1%
Accuracy: ±2% full scale
CO2 Resolution: 5ppm
Accuracy: ±(2% of range or 2% of reading)
1
CO Resolution: 1ppm
Accuracy: ±5% of range at constant
temperature and pressure
Display:
Liquid crystal display, 128 x 64 pixel
with RGB LED back-light
Controls:
Resin capped silicone rubber keys
Warm-up Time:
< 60 seconds
Response Time:
< 40 seconds N
2
0
< 60 seconds O
2
< 60 seconds CO
2
Operating Temperature:
0 – 40
o
C
Sensor Life:
2 - 3 years expected
6 month guarantee
Power Supply:
Mains power lead, USB connector,
Internal rechargeable Li-Ion battery
Time to Full Charge:
Approximately 3 hours.
Dimensions:
L 165mm, W 100mm, D 55mm
Weight:
500 grams
Connectivity: USB
Extra Features:
Data logging Capability
Visual and audible alarm
1
Conditions during factory calibration, typically 20°C, 1,000 mBar
Содержание G210