background image

  

  

  

3  

  

9.

 

When  the  test  is  finished  

  

  

  will  appear  at  the  bottom  of  the  screen.  

  

10.

 

If  a  high  reading  has  been  recorded,  you  can  mute  the  sound  by  pressing  

.  

  

11.

 

To  repeat  breath  test,  press  

  once  to  return  to  the  home  screen  and  repeat  steps  3-­‐8.  

  

12.

 

To  save  the  reading  press  

  and  select  the  relevant  patient  profile.  

  

13.

 

Remove  the  D-­‐piece  between  tests  to  purge  sensor  with  fresh  air.  

  

14.

 

To  switch  off,  press  and  hold  the  power  button  for  3  seconds,  unit  will  also  power  off  after  2  
minutes  of  inactivity  to  save  power.    
  

and  then  

  the  history  will  then  be  shown.  

  

Changing  date  and  time  

  

Press  

  and  then  

  you  can  then  select  either  d-­‐m-­‐y  or  m-­‐d-­‐y  for  the  date  format  and  12h  or  

24h  for  the  time  format.  Dates  and  times  are  then  adjusted  by  selecting  the  number  you  wish  to  

change  and  pressing  

or  

  to  increase  and  decrease.  Press  

  to  save  your  settings.  

  

  

Converting  %COHb  to  %FCOHb  –  

    

Reviewing  history  

  
The  Micro

+

basic

™  will  record  every  reading  taken  in  its  history  up  to  150  readings,  to  access  this  press     

Refer to the included Maternity Chart to convert the pregant mother's %COHb to 
%FCOHb.

Содержание coVita Micro+basic Smokerlyzer

Страница 1: ...Smokerlyzer Scientific contributions to health 0086 Former piCO For use with Micro basic www covita net 800 707 5751 ...

Страница 2: ...d Device Symbols 4 Environment 5 Instrument Layout 5 Display and Device Symbols 6 Warnings and Maintenance 8 Cleaning 8 Routine Maintenance 9 Troubleshooting 9 Calibration Procedure 10 COdata Software connecting to the PC 10 Returns Procedure 11 Spares 11 Warranty 11 2 Converting COHb to FCOHb 3 Reviewing history 3 Technical Specification ...

Страница 3: ...rbon monoxide is measured in parts per million ppm and blood carboxyhaemoglobin is measured in percentages COHb and fCOHb The three are compatible and convertible the ppm CO reading relates to gas in the lungs and on the breath i e the amount of poisonous CO that has been inhaled and the COHb and fCOHb readings relate to the percentage of vital oxygen that has been replaced in the bloodstream The ...

Страница 4: ...conds unit will also power off after 2 minutes of inactivity to save power and then the history will then be shown Changing date and time Press and then you can then select either d m y or m d y for the date format and 12h or 24h for the time format Dates and times are then adjusted by selecting the number you wish to change and pressing or to increase and decrease Press to save your settings Conv...

Страница 5: ...t Storage Pressure 800 1200mbar Operating Humidity 15 90 non condensing storage 0 95 Sensor Operating Life 5 years 5 year warranty Sensor Sensitivity 1ppm Sensor Drift 5 per annum Dimensions Approx 34 x 75 x 140 mm Weight Approx 200g including batteries Monitor Construction Case Polycarbonate ABS blend Micro basic Concentration Range CO 0 150ppm Display Full color touchscreen Detection Principle E...

Страница 6: ...ith COdata Software 7 Single use SteriBreath mouthpiece 8 Battery compartment 9 Battery compartment clip 10 Reset button Environment The Micro basic Smokerlyzer complies with the directive EN60601 1 2 electromagnetic compatibility but can be affected by cellular phones and by electromagnetic interference exceeding the levels specified in EN50082 1 This equipment should be moved if n ecessary to av...

Страница 7: ...pty Breath Test Environmental Test Settings Patient Profiles View patient results Inhale Hold Breath Countdown Timer Exhale ppm Reading COHb Reading Show fCOHb fCOHb Reading Display and Device Symbols Description Micro basic Battery condition Full ...

Страница 8: ...Traffic Light Cut off Levels Save Change Breath Hold Time History Mute Change date time Next Step Countdown to Sensor Change Change Sensor Calibrate Device Attach Flow Meter to Gas Canister Description Micro basic Ambient CO reading ...

Страница 9: ...own for 30 seconds will perform complete device reset this will clear any saved data and revert all settings to the factory defaults After performing a reset the device will need to have the date time set and be calibrated before it can be used Cleaning 1 The Smokerlyzer products are moulded with SteriTouch technology for optimum infection control and bacterial efficacy However Bedfont recommend w...

Страница 10: ...t on a non smoker gives an indication of CO being present it may indicate the following 1 A high ambient level of CO possibly due to a faulty combustion appliance or exhaust fumes 2 The effects of passive smoking 3 The device is contaminated by alcohol or another organic substance 4 The patient may have a very high level of hydrogen on their breath which can be generated in the human digestive sys...

Страница 11: ... lu b e h t e t a r lib a c o t w lo o o t is r o it n o m e h t f o e r u t a r e p m e t e h t If 4 r e i l p p su l ca o l r u yo r o t n o f d e B ct a t n co e s a le p s t is s r e p m le b o r p is h t if p u d e m ar w as h t i n u e h t e c n o r e at l n ai ag y Tr n w o sh e ill b w r e t e m o m r e h t d e r e h t e t a r lib a c o t h ig h o o t is r o it n o m e h t f o e r u t a r ...

Страница 12: ...or loss or damage in transit When your goods are received you will be se nt an email stating so 6 If you choose not to proceed with the repair a handling fee will be charged Ensure that you return the completed authorization form with an Official Purchase Order Number 7 The equipment will be returned to you as soon as coVita has received all relevant paperwork A carriage fee will be charged if the...

Страница 13: ...net 800 707 5751 coVita Tel 800 707 5751 Fax 800 721 2377 Email service covita net www covita net 212 Cottage Grove Ave Suite C Santa Barbara CA 93101 Issue 3 March 2017 Part No LAB679_USA Bedfont Scientific Limited reserves the right to change or update this literature without prior notice Registered in England and Wales Registered No 1289798 ...

Отзывы: