![Beckman Coulter SX4400 Скачать руководство пользователя страница 19](http://html.mh-extra.com/html/beckman-coulter/sx4400/sx4400_instructions-for-use-manual_2719762019.webp)
B01198AB
9
SX4400 Swinging Bucket Rotor
Preparation and Use
Figure 4
Seating Buckets on Yoke Pivot Pins
NOTE
Buckets must be inserted at an angle, or the bucket bottom will rest on the rotor base and the
pivot pins will not seat in the pin pockets.
Operation
1
Refer to the centrifuge instruction manual for centrifuge operation.
2
See
Run Speeds
, page 10, for information about speed limitations.
Removal and Sample Recovery
CAUTION
If disassembly reveals evidence of leakage, you should assume that some fluid
escaped the rotor. Apply appropriate decontamination procedures to the
centrifuge and accessories.
1
Remove the buckets from the centrifuge.
2
Remove covers (if used) from buckets, and remove the tubes from the buckets.
NOTE
If glass tubes break, remove the glass very carefully from the adapter and bucket. If all the glass
particles are not contained in the bucket or adapter, be careful when examining or cleaning the
centrifuge gasket and chamber as glass particles may be embedded in their surfaces.
1.
CORRECT
2.
INCORRECT
3.
Pivot pins are seated in bucket pockets
4.
Pivot pins are not seated in bucket pockets
Pivot pins are
seated in
bucket pockets
Pivot pins are
not seated in
bucket pockets
CORRECT INCORRECT
2
1
4
3
Содержание SX4400
Страница 4: ...B01198AB iv Revision History ...
Страница 9: ...ix Tables Tables 1 Available Tube Adapters for the SX4400 Rotor 3 ...
Страница 10: ...x Tables ...
Страница 27: ......
Страница 28: ......
Страница 30: ...B01198AB Warranty 2 Beckman Coulter Inc Benchtop Rotor Warranty ...
Страница 31: ......