![Beckman Coulter SX4400 Скачать руководство пользователя страница 16](http://html.mh-extra.com/html/beckman-coulter/sx4400/sx4400_instructions-for-use-manual_2719762016.webp)
B01198AB
6
SX4400 Swinging Bucket Rotor
Preparation and Use
Installing the Rotor Yoke
NOTE
Before the first use of a new rotor, lubricate the pivot pins with Spinkote (306812). After the first use,
apply Spinkote to the pivot pins approximately once a week and after cleaning or autoclaving.
NOTE
Refer to the centrifuge manual for proper installation and removal instructions. Failure to follow printed
instructions may result in damage to the rotor and the centrifuge.
1
Before installing the yoke in the centrifuge, lightly coat the drive hole with lubricant such as
Spinkote (306812) (see instructions under
Maintenance
).
2
Carefully lower the yoke straight down onto the centrifuge drive shaft (see
Figure 2
).
•
Be sure the yoke is properly seated on the shaft.
Figure 2
Installing the Rotor Yoke on the Centrifuge Drive Shaft
3
Fasten the tie-down screw (361367) onto the drive shaft.
•
Use the T-handle rotor wrench (365636) to tighten the tie-down screw firmly on the shaft.
CAUTION
Never drop the rotor yoke onto the centrifuge drive shaft. The drive shaft can be
damaged if the rotor is forced sideways or dropped onto it.
NOTE
If the yoke is left in the centrifuge between runs, before each run make sure it is properly seated on
the drive shaft and that the tie-down screw is tight.
1.
Lower the yoke onto the drive shaft
2.
Tighten the tie-down screw with the T-handle wrench
1
2
Содержание SX4400
Страница 4: ...B01198AB iv Revision History ...
Страница 9: ...ix Tables Tables 1 Available Tube Adapters for the SX4400 Rotor 3 ...
Страница 10: ...x Tables ...
Страница 27: ......
Страница 28: ......
Страница 30: ...B01198AB Warranty 2 Beckman Coulter Inc Benchtop Rotor Warranty ...
Страница 31: ......