PN B10093AE
1-5
Description
Controls and Indicators
1
Control Panel
The control panel (
) is mounted at an angle on the centrifuge top rear for easy visibility
and access. It contains a separate Actual Value Display and a Setup Screen, numerical keypad, and
touch keys for parameter entry and system control.
Figure 1.2
The Control Panel
Actual Value Display
) indicates the actual centrifuge operating conditions.
Figure 1.3
The Actual Value Display
1.
Actual Value Display
2.
Setup Screen
ROTOR ID
Indicates the installed rotor. At the start of a run, this field is blank. During acceleration,
the centrifuge identifies the installed rotor and displays the rotor name.
SPEED
Indicates rotor speed in revolutions per minute (RPM), or in relative centrifugal field, a
term describing the ratio of the centrifugal acceleration at a specified radius and speed to
the standard acceleration of gravity (RCF
g
).
1
2
7
8
9
4
5
6
1
2
3
0
E
N
T
E
R
CE
ROTOR ID
SPEED
TEMP
RPM/RCF
TIME
HR:MIN
START
STOP
IRIS
RPM
HOLD
EXP
g-sec
rad /sec
RCG (xg)
°c
ROTOR
SPEED
TIME
TEMP
A/D
OPT
ROTOR ID
SPEED
TEMP
RPM/RCF
TIME
HR:MIN
IRIS
RPM
HOLD
EXP
g-sec
rad /sec
RCG (xg)
°c
Содержание Avanti J-26S XPI
Страница 4: ...PN B10093AE iv Revision History ...
Страница 12: ...xii Contents Avanti J Series Centrifuge Warranty Related Documents ...
Страница 14: ...xiv Illustrations ...
Страница 16: ...xvi Tables ...
Страница 20: ...PN B10093AE xx Introduction Multi Compliance Label ...
Страница 36: ...PN B10093AE 1 16 Description Available Rotors ...
Страница 66: ...PN B10093AE 3 10 Troubleshooting JCF Z Rotor Identification ...
Страница 82: ...PN B10093AE A 10 Preinstallation Requirements Using J2 Series Rotors in the Avanti J 26S XPI ...
Страница 86: ...PN B10093AE Warranty 2 Beckman Coulter Inc Avanti J Series Centrifuge Warranty ...
Страница 87: ......