PN B10093AE
1-4
Description
Name Rating Plate
•
The user specifies whether the run should shut down or continue if the overtemp setting is
reached.
•
The system will always shut down using maximum brake if the temperature goes above 50°C.
Name Rating Plate
The name rating plate is affixed to the rear of the centrifuge. Check that the line voltage agrees with
the voltage listed on this name rating plate before connecting the centrifuge. Always mention the
model number and serial number when corresponding with Beckman Coulter regarding your
centrifuge.
Controls and Indicators
Key and Power Switches
The key switch and the power switch are located below the control panel (see
Figure 1.1
The Power Switch
The power switch is a two-position rocker switch (
I
, on;
O
, off) which controls electrical power to
the centrifuge.
The key switch has five positions, indicating the five different operating modes.
•
NORMAL
mode is used for regular, closed-door centrifugation.
•
ZONAL
mode is used when the JCF-Z continuous flow/zonal rotor is installed. In
ZONAL
mode,
the centrifuge can run up to 3000 RPM with the door open so that sample can be loaded and
unloaded while the rotor is spinning.
•
PROG(RAM) LOCK
mode disables the function that saves parameters to stored programs, to
prevent overwriting of previously stored programs.
•
RUN LOCK
mode disables all user inputs except
START
and
STOP
to ensure that current settings
do not get changed.
•
INST(RUMENT) LOCK
mode disables all user inputs except
STOP,
to prevent use of the
instrument after the current run is complete.
Содержание Avanti J-26S XPI
Страница 4: ...PN B10093AE iv Revision History ...
Страница 12: ...xii Contents Avanti J Series Centrifuge Warranty Related Documents ...
Страница 14: ...xiv Illustrations ...
Страница 16: ...xvi Tables ...
Страница 20: ...PN B10093AE xx Introduction Multi Compliance Label ...
Страница 36: ...PN B10093AE 1 16 Description Available Rotors ...
Страница 66: ...PN B10093AE 3 10 Troubleshooting JCF Z Rotor Identification ...
Страница 82: ...PN B10093AE A 10 Preinstallation Requirements Using J2 Series Rotors in the Avanti J 26S XPI ...
Страница 86: ...PN B10093AE Warranty 2 Beckman Coulter Inc Avanti J Series Centrifuge Warranty ...
Страница 87: ......