J326XP-IM-5AB
2-5
Description
Controls and Indicators
2
•
When
HOLD
is selected (no run time is specified), the
TIME
display shows the
time elapsed since
START
was pressed
. After 3 hours (180 minutes) have elapsed, the timer will reset to 0 and
continue counting elapsed time until
STOP
is pressed.
Knobs and Touch Keys
The rotor name, speed, time, and temperature settings are entered by turning the appropriate knob
until the required value is displayed.
Touch keys allow the user to start and stop the centrifuge, and to select speed mode (RPM or RCF),
time mode (specific run length or hold for continuous operation), and acceleration and deceleration
rates, and to toggle the display between set and actual values.
ROTOR
Knob
Used to select the rotor in use. As the knob is turned, the name of each
Beckman Coulter rotor that can be run in the centrifuge appears in succession in
the
ROTOR ID
display, and the LED by each rotor type abbreviation
(
JCF:
continuous flow rotors;
JSP:
special type rotors;
JS:
swinging bucket rotors;
JA:
fixed angle rotors;
JLA:
lightweight fixed angle rotors;
JV:
vertical tube rotors;
JE:
elutriator rotors) lights when appropriate.
SPEED
Selection
The
RPM/RCF
key is pressed to select the speed mode (RPM or RCF). After the
mode is selected, the
SPEED
knob functions as follows.
•
In
RPM
mode, each
SPEED
knob increment is 100 rpm at speeds up to
10 000 rpm. Above 10 000 rpm, each knob increment is 500 rpm.
•
In
RCF
mode, at settings less than or equal to 3000
g
, each
SPEED
knob
increment is equivalent to incrementing/decrementing the speed by 100
g
.
At settings greater than 3000
g
, each
SPEED
knob increment is equivalent to
incrementing/decrementing the speed by 250
g
.
•
Speed increments for elutriator rotors are either 10 rpm or 10
g
.
ROTOR
JSP
JS
JA
JLA
JV
JE
JCF
Содержание Avanti J-26 XP Series
Страница 6: ...J326XP IM 5AB vi Safety Notice Chemical and Biological Safety ...
Страница 12: ...xii Tables ...
Страница 16: ...J326XP IM 5AB xvi Introduction Recycling Label ...
Страница 32: ...J326XP IM 5AB 1 16 Specifications and Preinstallation Requirements Using J2 Series Rotors in the Avanti J ...
Страница 40: ...J326XP IM 5AB 2 8 Description Controls and Indicators ...
Страница 54: ...J326XP IM 5AB 3 14 Operation Unloading the Centrifuge ...
Страница 74: ...J326XP IM 5AB Warranty 2 Beckman Coulter Inc Avanti J Series Centrifuge Warranty ...
Страница 76: ... 2012 Beckman Coulter Inc All Rights Reserved www beckmancoulter com ...