TEMPERATURE KNOB ADJUSTMENT (Refrigerated Models Only)
If set and indicated temperature are not in agreement, the TEMPERA-
TURE knob may need adjustment.
Action
Result
1. Turn the centrifuge
power off.
2. Tap lightly on the °C meter
until the needle is centered
at 15°C.
°
C E N T I G R A D E
-10
0
10
20
30
40
3. Turn the power back on and
set the TEMPERATURE
control knob to 40°C.
4. Check that the door is
completely closed and the
manual latch lever is in
LOCK position.
5. Let the centrifuge run for a
short time until the tempera-
ture needle stabilizes.
°
C E N T I G R A D E
-10
0
10
20
30
40
6. Rotate the knob slowly
toward a colder setting (such
as 20°C) until the compres-
sor turns on. Loosen the two
setscrews in the knob (one on
each end of the knob) and set
the knob to the indicated
temperature.