3-3
Operation
Action
Result
3. Move the manual lock lever
to the left (UNLOCK).
UNLOCK
LOCK
4. Press DOOR to OPEN posi-
tion, then lift the door up.
Door remains open.
!
CAUTION
Never drop the rotor onto the centrifuge drive
shaft. The shaft can be damaged if the rotor is
dropped onto it or forced sideways.
5. Install the rotor or rotor yoke.
Be sure the base is seated on
the drive shaft, with the pin
in the rotor groove.
Drive Shaft
Pin
Groove
The rotor is ready to load. (You can load fixed angle rotors either
before or after installing the rotor in the centrifuge.)
6. Load the rotor according to
the rotor manual. Always run
the rotor with a balanced
load; fill all positions on a
swinging bucket rotor yoke
with buckets and/or carriers.
!
WARNING
Do not substitute a metal fastener for the
plastic tie-down nut supplied with the rotor.
If a tie-down nut comes off the shaft during
centrifugation, the
plastic
nut furnished will
break apart in the chamber, causing minimal
damage to the centrifuge. A loose
metal
object
could substantially damage the rotor,
chamber, and lid, and could potentially
escape the chamber into the laboratory,
causing personal injury or property damage.
Содержание Allegra 6 Series
Страница 1: ...BE IM 9 Allegra 6 Series and Spinchron R Centrifuges Instruction Manual...
Страница 6: ......
Страница 14: ...Allegra 6 Allegra 6R or Spinchron R Allegra 6KR 1 2 Description Figure 1 1 Allegra 6 Centrifuges...
Страница 38: ......
Страница 48: ......
Страница 49: ......
Страница 50: ......
Страница 52: ......