Generations/Simulations - 53
Current Generation (Source or Sink)
MC6-T's current generation can be done using one of two available
methods:
• MC6-T provides a 24 V loop supply voltage (source mode).
Setting:
Supply: On
.
• An external device provides the loop supply voltage (sink mode)
Setting:
Supply: Off
.
Connection depends on the loop supply setting. See
generation terminals. Internal supply. Range 0 … 55 mADC
Current generation terminals. External supply
Figure 38: Current generation terminals. Internal supply. Range 0 …
55 mADC
Figure 39: Current generation terminals. External supply
See also
Note:
If the connected instrument utilizes digital communication
and 24 V supply voltage is in use, the battery symbol is shown
in the user interface of
Documenting Calibrator
and
Data
Logger
.
Содержание MC6-T