44 - Active Terminals and Connections
Zeroing a Pressure Module
If the selected pressure module does not display zero gauge pressure
when the applied pressure is zero, the module has to be zeroed. To do so,
apply zero gauge pressure and tap the zero button.
Current Measurement
When measuring electric current, an important thing is selecting whether
MC6-T provides the 24 V
loop supply voltage
or not. If not, an external
device should provide the loop supply voltage.
Connection depends on the loop supply setting.
Figure 26: Current measurement terminals. Internal supply Range
-101 … +101 mADC
Figure 27: Current measurement terminals. External supply