WARNING:
Isolate power before starting any mainte-
nance procedures to prevent electrical shock
or accidental starting of equipment.
WARNING:
Pumps that have reached normal operating
temperature may have a high surface tem-
perature.
Do not perform any maintenance until after
a su
ffi
cient cool down period.
CAUTION:
Never perform any maintenance functions
while the unit is in operation.
Table 6.1 Maintenance Schedule
Item
Frequency
Action
Exhaust drip leg
Daily/Adjust as needed
Check for accumulated moisture
Oil / Oil
fi
lter (Synthetic)
2,000 hours
Change oil and replace oil
fi
lter
Exhaust
fi
lters
Check Monthly
Replace Annually
Check monthly and replace if required
Replace exhaust
fi
lters
Purge Filters
Check Monthly
Replace Annually or as needed
Inspect
Replace the inlet
fi
lter elements
Inlet
fi
lter screens / inlet
fi
l-
ters
Check Monthly
Replace Annually
Clear as required
Motor bearings*
1.5 - 10 HP motors
15 - 20 HP motors
25 HP motors
Annually
8 grams of grease per
fi
tting
17 grams of grease per
fi
tting
23 grams of grease per
fi
tting
Coupling
Every 2 years
Inspect coupling rubbers for wear.
Replace as needed.
1. Maintenance intervals depend on the pump’s use and ambient conditions. Monitor conditions of
oil regularly, and change when recommended or necessary.
2. (*) Motors have been lubricated at the factory. Motors that do not have the re-grease capability
(no grease
fi
ttings) are factory lubricated for the normal life of the bearings.
Refer to the service manual for grease speci
fi
cations.
6.0 Maintenance
6.1 General
Maintenance
Lubricated Rotary Vane Laboratory Vacuum Systems
4107 9021 88.01
6-1