Page 5
adequate ventilation to allow escape of gas in the event of
rupture of any of the gases.
Vertically mounted terminal units shall be mounted, when
facing the units, in the following order, horizontally from left
to right or vertically from top to bottom: -
O2, N2O, O2/N2O, Medical Air, Surgical Air, VAC, AGS.
When terminal units are mounted in a circular array on the
under surface of a pendant the sequence is as for vertically
mounted units but in a clockwise direction when viewed
from below. The AGS terminal unit may occupy the centre
of such an array.
Installation procedure
Terminal Unit. Fit. Fix to terminal unit mounting plate
contained within the pendant/boom with 2 - M4 bolts, nuts
and washers.
Hose Assembly. Connect.
Connect the gas specific NIST nut of each terminal unit to its
respective gas specific hose assembly.
Terminal unit. Commissioning.
Carry out the commissioning procedure when all terminal
units and hose assemblies have been fitted.
Installation of A Second Fix Assembly
General requirements.
The second fix assembly should only be fitted to the first fix
assembly when the wall surface, pendant, trunking or hose
boom installation is completed and clean conditions prevail.
Installation procedure
First fix assembly. Remove blanks. Remove dust plug or
pressure blank (if fitted) from the first fix assembly.
Valve. Assemble.
Locate spring into brass valve plug counterbore and
locate brass valve plug into front body. Check which thread
matches the thread on the first fix assembly, fit the ‘O’ Ring
to the base of this thread if not already present. Screw the
front body into the rear body (this is a left hand thread and
will tighten anticlockwise).
Tighten assembly with hand pressure only.
Terminal unit label. Fit.
Remove the backing paper from the terminal unit label.
Locate and affix to front body with self adhesive pad.
Terminal unit. Check operation.
Insert an AGS probe and check for correct operation.
Note...
Ensure that bezel is located with moulding marks next to the
wall to provide a neat installation.
Concealed installation. Locate bezel.
Locate bezel to inside lip of fascia plate.
Fascia plate. Fit.
Locate fascia plate squarely over the installation. With
surface installations, fascia lip fits securely inside plaster
box. Engage both screws with first fix gas specific bracket and
tighten.
Part No. 1827420-I3
© 2009 BeaconMedaes. All rights reserved.