Installation Instructions
Part number 1827420
Revision 3
07/10/09
Gem 10 AGS Terminal Unit
0088
Страница 1: ...Installation Instructions Part number 1827420 Revision 3 07 10 09 Installation Instructions Gem 10 AGS Terminal Unit 0088...
Страница 2: ...cified below Environmental Transport and Storage Conditions Min ambient temperature 0 degrees Celsius Max ambient temperature 40 degrees Celsius Min relative humidity non condensing 10 Max relative hu...
Страница 3: ...level to the centre line of the terminal unit Multi gas terminal unit installations when vertically mounted shall be spaced at 135 2 5mm between centres The centre of a vertically mounted terminal un...
Страница 4: ...t and mark copper stub pipe for length required Re move terminal unit assembly Copper stub pipe Cut to length Using a wheel cutter cut copper stub pipe to required length Ensure that ends are square f...
Страница 5: ...on is completed and clean conditions prevail Installation procedure First fix assembly Remove blanks Remove dust plug or pressure blank if fitted from the first fix assembly Valve Assemble Locate spri...