background image

BDDF00171 Issue 4

15/39

BD Alaris™ MRI Capsule

Features of the MRI Capsule

Power Input

The MRI Capsule is powered from the AC Power Supply through a standard IEC AC Power connector (i .e . C13, C14) . When 

connected to the AC Power Supply the AC Power indicator is illuminated . Both the Live and Neutral lines of the main 

supply are protected using fuses carried in a double fuse holder located on the AC Power Supply inlet connector .

When connected to the AC Power Supply, a three wire (Live, Neutral, Earth) supply must be used. If the 

integrity of the external protective conductor in the installation or its arrangement is in doubt, then the 

MRI Capsule must not be used.

w

To isolate the MRI Capsule from the AC Power Supply remove the AC Power cord from the source socket. 

The MRI Capsule should be positioned to allow access for disconnecting the AC Power cord.

Battery Supply

There are two batteries in the MRI Capsule:

MRI Capsule Internal Battery

The MRI Capsule should normally be operated from the AC Power supply . However, in the event of temporary loss of AC 

Power, an internal power supply will provide approximately 60 minutes of backup power . AC Power to the Pumps will be 

discontinued .

AC Power should be re-applied as soon as possible as there will be no AC Power supplied to the Pumps 

while the MRI Capsule is running on the battery. 

Note:  When the AC Power is disconnected, the Pumps on the MRI Capsule Medical Device Interface (MDI) tiles (herein 

after referred to as tiles) that generate an alarm will be replicated on the Pump Alarm Location Beacon .

The Battery indicator illuminates whenever the MRI Capsule is running from the internal battery . When illuminated the 

MRI Capsule is operating from internal battery; when flashing the battery power is low and auto power down is imminent . 

The battery is automatically charged whenever the MRI Capsule is connected to the AC Power Supply . As the MRI Capsule 

is designed to operate from the AC Power supply it will only power up when connected to the AC Power Supply .

w

If transfer of the MRI Capsule is required, then prior to disconnecting from AC Power supply the User 

must ensure there is sufficient battery power on each of the Pumps.

w

In the event of a power loss, the MRI Capsule will provide a visual indicator and emit an audible tone 

every 30 seconds for the first 14 minutes to alert operators. This audible tone and visual indicator 

will escalate to every 15 seconds after 14 minutes until the battery is fully depleted. These tones and 

indications should not be confused with the continuous alarm and LED notification that is initiated if the 

MRI Capsule exhibits a fault condition.

Magnetic Indicator Battery

The Magnetic Indicator has an independent Primary Alkaline Battery . It is a non-rechargeable battery to provide power 

to the Magnetic Indicator . The function of the battery is to operate the Magnetic Indicator independently from the AC 

Power supply or the MRI Capsule internal battery . The battery indicator of the Magnetic Indicator will be flashing when 

the battery power is low and BD recommends replacing the battery every two years . Please contact Qualified Service 

Personnel to replace the battery .

Note:  The Magnetic Indicator will always be on .

AC Power Output to Infusion Pumps

The MRI Capsule has its own power distribution circuit to supply AC Power to the attached Pumps . Each MRI Capsule tile 

IEC connector will have AC Power present on the connector, without the Pump attached to the MRI Capsule, and the tile 

green LED will be illuminated . When a Pump is fully attached to the MRI Capsule tile the AC Power indicator on the Pump 

will illuminate to indicate the Pump is powered and charging .

The MRI Capsule tile AC Power outlet connection is intended only for connection to the aforementioned 

Pumps. Please refer to the compatibility matrix in the ‘Intended Use’ section of this guide. Never attach 

any other equipment to the outlet connector.

Содержание Alaris 80300MRI01-33

Страница 1: ...BD Alaris MRI Capsule Model 80300MRI01 33 en Directions For Use s...

Страница 2: ...Power Input 15 Battery Supply 15 AC Power Output to Infusion Pumps 15 System Notifications 16 System Fault Indication 17 Pump Alarm Location Beacon 18 Operating Precautions 19 Operating Environment 19...

Страница 3: ...DDF00171 Issue 4 2 39 BD Alaris MRI Capsule Spare Parts 33 Spare Parts 33 Contact Us 34 Document History 34 Electromagnetic Emissions 35 Electromagnetic Immunity 36 Recommended Separation Distances 38...

Страница 4: ...l communicate with the Pump Alarm Location Beacon for High Medium and Low priority alarms where applicable For further alarm information refer to Pump DFU Compatible Alaris Infusion Pumps for MRI Caps...

Страница 5: ...ts compatible Pumps can only be used by trained and qualified medical staff within the MRI environment Intended patient population The MRI Capsule can only be used on one patient at a time and it serv...

Страница 6: ...this manual For example see Battery Supply Italics Used to refer to other documents or manuals For example refer to Pump DFU Also used to define special terms e g MRI Capsule Note Notes contain suppl...

Страница 7: ...rnal battery when flashing the battery power is low and auto power down is imminent AC Power Indicator When illuminated the MRI Capsule is connected to the AC Power supply and the battery is being cha...

Страница 8: ...um Yellow LED flashes Yes Move the MRI Capsule until the Magnetic Indicator green LED flashes Magnetic field density 40mT High Red LED flashes Yes Move the MRI Capsule until the Magnetic Indicator gre...

Страница 9: ...irements of Council Directive 93 42 EEC as amended by 2007 47 EC IPX2 Protected against water droplets deflected up to 15 from vertical T Date of Manufacture t Manufacturer U Not for Municipal Waste M...

Страница 10: ...with bag hanger hooks Pump Alarm Location Beacon Keypad with Indicators Door handle with a lock Grooves for Infusion Sets from Infusion Pumps Grooves for Infusion Sets to Volumetric Pumps Manual Brak...

Страница 11: ...with the MRI Capsule handle Squeeze and hold the brake lever to release the brakes and locate the MRI Capsule in the desired position To engage the automatic brakes release the brake lever Manual Brak...

Страница 12: ...BDDF00171 Issue 4 11 39 BD Alaris MRI Capsule Features of the MRI Capsule Configurations 3 Volumetric and 3 Syringe Pumps 4 Volumetric Pumps 4 Syringe Pumps...

Страница 13: ...e Free Valve Auxilary set 205cm 1 spiked drip chamber Hydrophobic air inlet filter 20 drops per ml 2 pinch clamps Pressure sensor disc 15 m Drip Chamber Filter Alaris GH Syringe Pump Alaris PK Syringe...

Страница 14: ...en closing the MRI Capsule door and when installing the pumps After installing a pump ensure its related set is correctly placed within the grooves of its associated outlet w All these sets and access...

Страница 15: ...tainer Raise the fluid container higher than the minimum height where possible to minimise loops and kinks in the infusion set tubing 45cm minimum To operate the clamp 1 Grip the handle at the lower e...

Страница 16: ...ted to the AC Power Supply As the MRI Capsule is designed to operate from the AC Power supply it will only power up when connected to the AC Power Supply w If transfer of the MRI Capsule is required t...

Страница 17: ...nly System Notifications that use the primary speaker System Fault Indications are represented by the System Fault Indicator and use the secondary speaker The Pump Alarm Location Beacon is used to rep...

Страница 18: ...es the System Fault Indicator and activates the audible tone each time the MRI Capsule is switched on Trigger Visual Indicator Audio Indicator Description Communication Failure System Fault Indicator...

Страница 19: ...l Indicators can be found below Pump Alarm Priority Colour Frequency HIGH Priority Red Flashing faster than Medium Priority MEDIUM Priority Yellow Flashing or solid LOW Priority Yellow Solid Note In t...

Страница 20: ...Ib device When Alaris Infusion Pumps are attached and operational the system becomes a Class IIb system This MRI Capsule is a Class IIb device Therefore its RF emissions are very low and are not likel...

Страница 21: ...d exists if the MRI Capsule s casing is opened or removed Refer all servicing to qualified service personnel When connected to an external power source a three wire Live Neutral Earth supply must be u...

Страница 22: ...e the Magnetic Indicator flashes green The infusions can continue without interruption The Pumps must be checked by qualified service personnel when the infusions are complete If the MRI Capsule is mo...

Страница 23: ...ome fully operational w Do not switch off the MRI Capsule during this initial 90 second period w If any of the verification checks fail when the MRI Capsule is switched on suspect a fault Remove the M...

Страница 24: ...position when fitted to the bar Verify that the AC Power indicator on the Pump is illuminated when the MRI Capsule is connected to AC Power 7 Verify Pump is secure by gently pulling the Pump away fro...

Страница 25: ...Capsule Note When removing an Alaris VP Plus Guardrails Volumetric Pump from the top left MRI Tile follow the best practice specified below w If the indicator in the MRI Capsule tile is not illuminat...

Страница 26: ...rake lever and apply the manual brakes After placing the MRI Capsule in position you must ensure that all the braking systems are activated and working correctly Inadvertent movement of the MRI Capsul...

Страница 27: ...sla Static Magnetic field of MRI Scanner 1 5 tesla 200mT 2000 gauss 0 6m 0 5m No 70mT 700 gauss 0 9m 0 8m No 40mT 400 gauss 1 1m 1 0m No 30mT 300 gauss 1 2m 1 1m No 20mT 200 gauss 1 4m 1 2m Yes 10mT 1...

Страница 28: ...d up to 15 from vertical Battery MRI Capsule Battery Type Nickel Metal Hydride Charge time With no Pumps attached Up to 8 hours to 95 charge time based on battery life cycle With Pumps attached Up to...

Страница 29: ...irect connection between the MRI Capsule and the potential equalisation busbar of the electrical installation To use the Potential Equalisation Connector connect the Potential Equalisation Connector o...

Страница 30: ...ed periods of storage At least once per year Refer to TSM for identification of parts Inspect AC Power outlets and the AC Power inlet for damage Perform electrical safety checks The complete unit leak...

Страница 31: ...to the compatible cleaning agents for The MRI Capsule exterior The MRI Capsule plastic interior sides and the door The following products were tested and are acceptable for use on the MRI Capsule if u...

Страница 32: ...BDDF00171 Issue 4 31 39 BD Alaris MRI Capsule Maintenance...

Страница 33: ...tion Disposing of this product correctly will help to save valuable resources and prevent any potential negative effects on human health and the environment which could otherwise arise from inappropri...

Страница 34: ...his MRI Capsule is included within the Technical Service Manual The Technical Service Manual is now available in electronic format on the World Wide Web at bd com int Alaris technical A username and p...

Страница 35: ...bd com New Zealand Freephone 0800 572 468 NZ_customerservice bd com Norge 47 64 00 99 00 bdnorge bd com Polska 48 22 545 05 80 dok medicart eu Portugal 351 219 152 593 dl pt geral carefusion com South...

Страница 36: ...ons Group 1 The MRI Capsule uses RF energy only for its internal function in the normal product offering Therefore its RF emissions are very low and are not likely to cause any interface in nearby ele...

Страница 37: ...ound for Input Power Ports 2 kV Line to ground for Signal input output 0 5 kV 1 kV Line to line for Input Power Ports 0 5 kV 1 kV 2 kV Line to Ground for Input Power Ports 2 kV Line to ground for Sign...

Страница 38: ...tic site survey b should be less than the compliance level in each frequency range c Interference may occur in the vicinity of equipment marked with the following symbol Note 1 At 80 MHz and 800 MHz t...

Страница 39: ...30 10 1 11 3 80 3 80 7 28 100 3 50 12 00 12 00 23 00 For transmitters rated at a maximum output power not listed above the recommended separation distance d in metres m can be determined using the eq...

Страница 40: ...I Capsule should be observed to verify normal operation in the configuration in which it will be used Test specifications for ENCLOSURE PORT IMMUNITY to RF wireless communications equipment in accorda...

Страница 41: ...o Alaris and Guardrails are trademarks of Becton Dickinson and Company or its affiliates 2019 BD All rights reserved BD Switzerland S rl Route de Crassier 17 Business Park Terre Bonne Batiment A4 1262...

Отзывы: