Chapter 2: Routine Maintenance
Advisor
®
Service Manual
2-5
Co
2
HILo Calibration
Figure 2.2: HILO CAL
noTE! A HILo CAL procedure should be performed after the
capnography parameter has been on for at least 15
minutes.
noTE! Remove the device from the patient before performing a
high/low calibration procedure.
To perform a HILO CAL:
If the patient is attached to the capnography module, disconnect
1.
the sample line tubing from the monitor by unscrewing the filter
luer from the module on the left side of the monitor.
If you are using a standard pneumatics module, remove the
2.
moisture filter and attach the calibration adapter (BCI
®
part
number 8223) to the gas inlet.
Connect one end of the “T” assembly to the regulator on the gas
3.
canister. Connect the second end of the “T” assembly to the gas
inlet on the monitor. Leave the third end of the “T” assembly open
to room air.
Turn the rotary knob on the monitor to move the cursor. Highlight
4.
the CO
2
parameter box name and push the knob to select.
Highlight HILO CAL and push the knob to select.
5.
Turn the rotary knob to highlight YES and push the knob to select.
6.
TURN GAS ON will be displayed in the CO
2
parameter box.
Quickly open the flow control valve on the calibration gas canister. The valve must be fully opened in less
7.
than 30 seconds.
When TURN GAS OFF is displayed in the CO
8.
2
parameter box, close the flow control valve of the calibration
gas canister.
When the calibration procedure is finished, CAL DONE will be displayed in the CO
9.
2
parameter box.
Change the units of measurement for CO
10.
2
to percent concentration (%) as follows.
Turn the rotary knob on the monitor to move the cursor. On the main menu at the bottom of the display,
a.
highlight
SETUP
and push the knob to select.
Highlight
b.
PARAMETER OPTIONS
and push the knob to access the parameter options submenu.
Highlight
c.
CO2 UNITS
and push the knob to select.
Highlight the desired unit and push to select.
d.
Highlight
e.
MAIN
or
PREVIOUS
and push the knob to select.
Verify the accuracy of the calibration by opening and closing the flow control valve on the calibration gas
11.
canister at a rate approximately 15 Hz (on for two seconds, off for two seconds). Repeat this for 4-8 on/off
cycles.
Verify that the ETCO
12.
2
reading in the CO
2
parameter box indicates 10.0% CO
2
0.4 (9.6-10.4%)
Disconnect the calibration gas fixture. If you are using a standard pneumatics module, remove the
13.
calibration adapter and reconnect the moisture filter.
If the calibration procedure is unsuccessful, a message will be displayed in the CO
•
2
parameter box. The
capnography parameter will revert to the last successful calibration data and resume operation.
4
1
2
3
Gas inlet
T connector
Calibration gas
Flow control valve
4
1
2
3
Содержание Advisor
Страница 2: ......
Страница 6: ...Table of Contents iv Advisor Service Manual This page is intentionally left blank ...
Страница 8: ...Revision History vi Advisor Service Manual This page is intentionally left blank ...
Страница 32: ...Chapter 2 Routine Maintenance 2 10 Advisor Service Manual This page is intentionally left blank ...
Страница 42: ...Chapter 3 Block Level Description 3 10 Advisor Service Manual This page is intentionally left blank ...
Страница 43: ...Chapter 4 Diagrams Advisor Service Manual 4 1 Chapter 4 Diagrams ...
Страница 44: ...Chapter 4 Diagrams 4 2 Advisor Service Manual ...
Страница 45: ...Chapter 4 Diagrams Advisor Service Manual 4 3 ...
Страница 46: ...Chapter 4 Diagrams 4 4 Advisor Service Manual ...
Страница 47: ...Chapter 4 Diagrams Advisor Service Manual 4 5 ...
Страница 48: ...Chapter 4 Diagrams 4 6 Advisor Service Manual ...
Страница 49: ...Chapter 4 Diagrams Advisor Service Manual 4 7 ...
Страница 50: ...Chapter 4 Diagrams 4 8 Advisor Service Manual ...
Страница 54: ...Chapter 5 ECG Technical Reference 5 4 Advisor Service Manual This page is intentionally left blank ...
Страница 56: ...Appendix App 2 Advisor Service Manual This page is intentionally left blank ...
Страница 57: ......