Quick
Start
COVIDAIR
S/T
Easy
English
1.1.docx
/
BCD
/
30.04.2020
COVIDAIR S/T Easy Quick Start Guide
Page
13
13
11. Setting the rise time in seconds. The value of the rise time is between 0.1 and 1 in 0.05s
increments. A short rise time (< 0.3 seconds) can lead to an overshoot).
12. Setting the drop time in seconds. The value of the descent time is between 0.1 and 1.45
in 0.05s increments.
13. Setting the humidity. The humidity value is between 1 and 5 in 1 increment. Do not use
this setting if the humidifier is not used.
14. Humidity activation slider (see its meaning and use at the end of the chapter)
For each setting instruction, the "-" button allows you to decrease the value while the "+"
button allows you to increase the value by the indicated steps. It is also possible to let your
finger pressed in order to reach the settings faster.
The sliders in front of certain parameters allow you to activate or not the taking into
account of the adjusted parameter. A simple press on the cursor changes its value.
When the cursor is ON, the parameter is active
When the cursor is OFF, the parameter is not active.
WARNING
If the parameter cursor is OFF, changing the parameter has no influence on the actual ventilation.