2.5.3. Ventilation Parameter Page
The parameters menu will determine the COVIDAIR
ventilation parameters. All parameters can
be changed during ventilation. This menu is only accessible with a password. The default
password 0000
WARNING
The adjustment and modification of the parameters must be carried out under the supervision of
a doctor or a member of the medical profession trained in ventilation.
1.
Setting the minimum respiratory rate per minute imposed on the patient. The value of
the respiratory rate is between 3 and 30 with steps of 1 breath per minute.
2. Adjustment of the EPAP expiratory pressure in cmH
2
O. The value of the expiratory
pressure is between 2 and 40 cmH
2
O it 1 cmH
2
O increments, but it cannot be greater than
the IPAP value. If IPAP and EPAP are equal the device functions as CPAP.
3. Adjustment of the IPAP inspiratory pressure in cmH
2
O.
4. Tidal volume adjustment. WARNING: This value did not take care of leaks. It is not
recommended to use this value for ventilation system with leak.
The tidal volume value is between 200 and 4000 milliliters with 50ml increments.
5. Tidal volume activation slider (see its meaning and use at the end of the chapter)
6. Respiratory Rate adjustment in %. The end of the inspiratory cycle is reached when the
flow rate drops to X % of the peak flow rate. This value can be set between 15% and 99%
in 1% increments.
7. Respiratory Rate cycling activation slider (see its meaning and use at the end of the
chapter)