background image

 

11 

 
 
Earth/Ground Leakage Current 

 
With the Earth/Ground function selected and the Ground-Open, the display will show the 
leakage current in the ground wire of the DUT.   The display is in µAmps and will read from 
0 to 1999. 
 
Selecting this function automatically opens the connection to Earth/Ground and passes any 
leakage current through a 1000 Ohms load with either AAMI ES1-1993 or IEC 601 
frequency compensation as selected by the Load key.   
 
The test should be performed for all receptacle switch combinations as called out in the 
Test Matrix, or a subset of these based on the manufacturer’s specifications or local codes 
and protocols. 
 
NOTE:  This test has no meaning for equipment that does not use a three-wire 
(Earth/Grounded) power cord. 
 

Enclosure Leakage 

 
With the Enclosure function selected, the display will show the leakage current between the 
Enclosure (Chassis) and Earth/Ground.  The display is in µAmps and will read from 0 to 
1999. 
 
The test requires that the Black (Kelvin) test cable be plugged into the Chassis Socket. The 
other end should be connected to a solid ground point on the DUT.  Consult the equipment 
manufacturer’s documentation for the recommended connection point. 
 
NOTE:  If a non-conductive enclosure is used, a 200 cm

2

 conductive foil pad should be 

used.  This foil is to be placed in close contact with the enclosure and connected to the 
Black lead. 
 
Any leakage current will flow through the Black lead and then through a 1000 Ohms load 
with either AAMI ES1-1993 or IEC 601 frequency compensation as selected by the Load 
key.   
 
The test should be performed for all receptacle switch combinations as called out in the 
Test Matrix, or a subset of these based on local codes, facilities practices and equipment 
manufacturers recommendations. 
 

Содержание SA-2001

Страница 1: ......

Страница 2: ...RESERVES THE RIGHT TO MAKE CHANGES TO ITS PRODUCTS OR SPECIFICATIONS AT ANY TIME WITHOUT NOTICE IN ORDER TO IMPROVE THE DESIGN OR PERFORMANCE AND TO SUPPLY THE BEST POSSIBLE PRODUCT THE INFORMATION IN...

Страница 3: ...ting the Device Under Test DUT to the Analyzer A serious hazard may occur if the patient is connected when testing with the Analyzer Do not connect any leads from the patient directly to the Analyzer...

Страница 4: ...S AAMI ES1 1993 or IEC 601 SELECTABLE TEST LOADS 85 TO 265 VAC OPERATION FULL 20 AMP RATING TOUCH CONTROL KEYS NO KNOBS LED STATUS INDICATORS AUDIO FEEDBACK SELF TEST POINTS EXTERNALLY REPLACEABLE GRO...

Страница 5: ...hat is being performed and the current setup This section will review the layout and operation of each of the elements LAYOUTS 10 Universal Patient Lead Inputs Unit indicators Volts Amps Amps Ohms Tes...

Страница 6: ...ture that first turns off the power to the DUT delays for a short period reverses the lines and the turns the unit back on This eliminates the need for the user to remember to delay at this point or r...

Страница 7: ...ocated on the face of the unit for ease of connection They accept both standard snap and banana jack connectors The spring loaded ball construction insures a good connection for the various sizes of b...

Страница 8: ...s and External Connectors 31 2 digit LCD Display ISO Test indicator and key Load Selection indicator and key AAMI IEC Receptacle for DUT power cord Power Cord for unit Function indicators with two lig...

Страница 9: ...l Connectors 31 2 digit LCD Display Load Selection indicator and key AAMI IEC Receptacle for DUT power cord Power Cord for unit Function indicators with two light touch arrow keys for selection Mains...

Страница 10: ...that the unit be returned to this condition when plugging and unplugging the Device Under Test DUT Power Receptacle Confirmation Once plugged in the first step is to insure that the wall receptacle t...

Страница 11: ...ill show the resistance between the Chassis Test lead and the Earth Ground Pin on the Analyzer Test Receptacle This resistance is a combination of the resistance within the DUT enclosure and the resis...

Страница 12: ...function selected the display will show the leakage current between the Enclosure Chassis and Earth Ground The display is in Amps and will read from 0 to 1999 The test requires that the Black Kelvin t...

Страница 13: ...es the leakage current that would flow through the leads if the patient were to come into contact with Earth Ground Lead to Lead Leakage With the Lead to Lead function selected the display will show t...

Страница 14: ...ound through a 1000 Ohms load with either AAMI ES1 1993 or IEC 601 frequency compensation as selected by the Load key The test should be performed for all receptacle switch combinations as called out...

Страница 15: ...rrent With the External function selected and the Isolation Voltage key depressed the display will show the isolation leakage current between the test leads The display is in Amps and will read from 0...

Страница 16: ...e Chassis Socket and the other in the External Socket The other ends should be connected to the points of interest Since the Black Kelvin cable removes any error from the test leads providing a true 4...

Страница 17: ...en Reverse Closed Closed On Open Reverse Closed Open On Open Reverse Open Closed On Open Reverse Open Open Off Closed Forward Closed Closed Off Closed Forward Closed Open Off Closed Forward Open Close...

Страница 18: ...LVE MONTHS FROM THE DATE OF SHIPMENT EXCLUSIONS THIS WARRANTY IS IN LIEU OF ANY OTHER WARRANTY EXPRESSED OR IMPLIED INCLUDING BUT NOT LIMITED TO ANY IMPLIED WARRANTY OF MERCHANTABILITY OR FITNESS FOR...

Страница 19: ...git 100 KHz 1 00 MHz 5 0 of reading 1 digit ISOLATION TEST 110 Line Voltage 5 LOAD AAMI ES1 1993 or IEC 601 Selectable CAPACITY 15 Amps 30 Minutes 20 Amps 5 Minutes PHYSICAL ENCLOSURE 8 625 x 5 5 x 1...

Страница 20: ...19 The following is a representation of the label found on the back of the SA 2010 APPENDIX A LABELS...

Страница 21: ...20 The following is a representation of the label found on the back of the SA 2005...

Страница 22: ...21 The following is a representation of the label found on the back of the SA 2001...

Страница 23: ...22 NOTES...

Отзывы: